aeroplus E
Taking into service
21
NOTE!
Only use the supplied administration accessories, such as nasal cannulas, tubes and
especially humidifiers.
The use of accessories not specified for use with the oxygen concentrator may affect
its performance.
The responsible organization is also responsible for the compatibility of all components
of the system. This has to be checked prior to the initial use of the systems.
7.2 Choosing the location
Please consider the following when choosing the location:
– The device should have 30cm clearance from any walls, curtains and other large
objects (e.g. cupboards), to ensure unrestricted entry of air through the back of the
device.
– The device is air cooled. It must therefore not be placed near heaters, etc. At such a
location there is a risk of overheating.
– The device should be operated only in places where it is not impaired by air pollution
or smoke.
– Always place the device on a horizontal surface.
NOTE!
The device can be safely lifted and moved by the carrier handle on top of the device.
ATTENTION! Ensure a sufficient air supply!
The aeroplus E oxygen concentrator is not to be used directly next to or stacked
with other equipment. Ensure sufficient distance to walls etc.!
Place the aeroplus E oxygen concentrator so that the air inlet and outlet are in a
well-ventilated area.
ATTENTION! Oxygen pollution!
Only install the oxygen concentrator where it is not affected by pollution in the
air, such as particles and dust, or by smoke! In general, the device should be
installed in a low-pollutant area.
ATTENTION! Humidity
Do not use or store to device in wet or moist areas!
Attention! Light and sunlight!
Only place the oxygen concentrator where it is not exposed to a direct light
source/heat source, such as direct sunlight!
Содержание aeroplus E
Страница 1: ...16 Instructions for Use aeroplus E ...
Страница 46: ...aeroplus E 46 ...