Alarm category
Description
Sensor
LED: permanently yellow
Ton: no Signal
Possible Cause:
The sensor signals have reached a level below 20%. The
oxygen concentration may not be exactly determined.
Countermeasures:
– Inform the customer service.
System
LED: All yellow flashing
Tone: Medium priority
Possible Cause:
Micro-processor fault.
Countermeasures
– Switch off the unit immediately and inform the customer
service.
NOTE!
Output and generation of oxygen is stopped for safety
reasons.
Gas temperature
LED: yellow flashing
Tone: Medium priority
Possible Cause:
The communication with the temperature sensor is interrupted.
The oxygen concentration may not determined.
Countermeasures:
– Inform the customer service.