![Krober aeroplus 5 Скачать руководство пользователя страница 71](http://html1.mh-extra.com/html/krober-0/aeroplus-5/aeroplus-5_instructions-for-use-manual_2004043071.webp)
aeroplus 5
Technical data
71
5 Technical data
Model
aeroplus 5
Classification acc. to MPG
IIa
Operating voltage
230 V, 50 Hz
Ambient operating temperature
Operation: +10 to +40 °C
Storage: -20 to +70 °C
Sound pressure level
< 40 dB(A)
Power input
295 W
Coarse dust filter
in back of housing
Inlet filter
in back of housing
Fuses
Internal: 1 x T3,15A H 250 V
internal: 1 x T1,0A L 250 V
Temperature range
+10 to +40 °C
Atmospheric pressure range
700 mbar to 1060 mbar
Weight
15 kg
Dimensions (HxWxD)
(60 x 29 x 40) cm
O2-concentration
1 to 3 l/min.
95 % - 3 %
3 to 4 l/min.
93 % +/- 3 %
4 to 5 l/min.
85 % +/- 3 %
Status display O2-concentration
(at operating temperature)
84% +/- 2 % status display
Low-level Priority Alarm
60% +/- 2 % deficiency display
Mid-level Priority Alarm
max. recommended volume flow
5 l/min
Volume flow (at 0 or 7kPa)
0.5 – 5 l/min acc. to pre-setting
max. output pressure:
70 kPa
Overpressure relief
200 kPa (oxygen reservoir)
360 kPa (compressor)