Alarm category
Description
O2-Concentration < 60%
LEDs: yellow flashing
Tone: Medium priority
Possible Cause:
The oxygen concentrator
aeroplus 5
is equipped with a sensor
to monitor the oxygen concentration of the oxygen output.
If this drops below 60%, the lack of oxygen alarm is triggered.
Countermeasures:
– Inform the customer service.
O2-Concentration < 82%
LEDs: permanently yellow
Tone: Low priority
Possible Cause:
The oxygen concentrator
aeroplus 5
is equipped with an
innovative multi-function sensor to monitor the oxygen
concentration of the oxygen output.
If this drops below 82% (but more than 60%), the oxygen status
alarm is triggered.
Countermeasures:
– Inform the customer service.
NOTE!
A low priority tone means NO acoustical signal.
Содержание aeroplus 5