![Krober aeroplus 5 Скачать руководство пользователя страница 84](http://html1.mh-extra.com/html/krober-0/aeroplus-5/aeroplus-5_instructions-for-use-manual_2004043084.webp)
aeroplus 5
Operation
84
ATTENTION!
The unit has been designed for operation at altitudes of up to 2000 m above seal
level. If the unit is used beyond this specification, compliance with the specified
performance data can no longer be guaranteed.
8.3 Alarms
WARNING! Risk of health damage!
If an alarm cannot be switched off by applying certain countermeasures, you
should immediately change over to an alternative oxygen supply (e.g. oxygen
cylinder). You should also immediately contact the customer service of the
manufacturer.
8.3.1 Alarm priorities
We differentiate between three alarm priorities:
Alarm
priority
Description
Acoustics
High
WARNING! Risk of health
damage!
Immediate countermeasures required to
save the patient from any harm.
dididi--didi----dididi-didi ---------
dididi--didi----dididi-didi---------
Medium Quick countermeasures by the user are
required.
di-di-di----------------------------------
-----------------------------------------
di-di-di
Low
Utmost attention of the user is required.
Without acoustic alarm
NOTE!
The alarm priorities can audibly be differentiated by different alarm sound sequences.
With a higher priority the number of alarm signals per unit of time increases.