27
Chapter 7:
Troubleshooting
7.1 Error Messages
The MetaCheck™ may detect certain errors and display them on the LCD of the
MetaCheck™. It will give the error number and a brief description of the
problem. Additional information is provided below.
7.1.1 Error 100: No Printer Detected
ERROR 100
No Printer Detected
[Results]
[Print]
Figure 7.1 Error 100 Screen
If this appears, the MetaCheck™ cannot detect a printer.
1. If this error appeared while you were printing, wait until that job is
finished, and then press “enter” again.
2. Verify that the printer cable is firmly plugged into each device.
3. Turn off each device. Remove the cable from each end, plug them
back in, and turn the power back on. The measured data from your
test will be saved until you are successfully able to print a report or
until a new test is executed.
4. Check to see if there is a paper jam in the printer that needs to be
cleared.
7.1.2 Error 101: System Timed out. No Breath Detected
ERROR 101
System Timed Out.
No Breath Detected.
[Restart]
Figure 7.2 Error 101 Screen