8
Chapter 3:
Quick Use Instructions
1. Turn on the unit.
CAUTION: Ensure the Single Patient Use
“MetaBreather™”
hose is not attached during
calibration. Remember to dispose of the hose/mouthpiece after use.
2. Wait for device to be ready.
o
Calibration will start automatically when device is turned on.
o
When calibration is complete the device indicates
“Waiting for First Breath
”.
3. Begin the test.
o
Extend the hose to full length.
o
Place the mouthpiece of the
NEW
Single Patient Use
“MetaBreather™”
in the
patient’s mouth.
o
Attach the hose to the
“Air Intake”
on the device.
o
Place the nose clip securely on the patient’s nose.
4. Perform the test.
o
Advise the patient that the lips need to create a seal around the mouthpiece
o
Encourage the patient to relax and breathe normally.
o
When test is complete the device indicates
“Test Complete”
“Remove the Hose”
.
o
Dispose of the
“MetaBreather™”
hose, mouthpiece and nose clip after use.
5. Enter the patient’s information (Y/N).
o
If comparison data and BMI calculation is desired, select “Yes” then use the
arrow keys to enter the patient’s age, weight, height, and gender.
o
Otherwise, select “No.”
6. Print the report.
o
The report can be printed after the patient’s information has been entered, by
pressing the “Print” soft key.
o
The machine will automatically begin the “Drying Mode” when ON/OFF key is
pressed. When performing successive tests this mode may be bypassed by
pressing the ON/OFF key twice.