454200/A
21
Calibration
Topics
First time roll and pitch calibration of a PX sensor, page 21
Manually entering the known pitch and roll calibration values of a sensor on the trawl
door, page 22
First time roll and pitch calibration of a PX
sensor
If you want your sensor to measure the pitch and roll of the trawl doors, you must
calibrate it.
Procedure
1
Place the sensor in the door adapter.
On the port door, the “down element shall face down”. On the starboard door, the
“side” element shall face down.
2
Place the door on a stable surface.
a
Use a level and set the door in correct upright position to create a 0 degrees
pitch and 0 degrees roll reference.
3
Connect the sensor to the computer using the programming cable.
4
On the processor unit desktop, double-click the relevant program icon to start it.
5
In the very first
PX Sensor Configurator
dialog box, select communication port
and sensor type, then select
Connect
.
The dedicated
Configurator
dialog box opens.
6
Select
Calibrate Sensor
to open the dialog box.
7
Provided that the door is in its leveled position, select
Leveled
..
8
Tilt the door at least 20 degrees inwards in the roll direction and make sure the
physical door pitch does not change.