Replacing the Lens
Replacing Procedure
The currently installed lens can be verified by the status information.
To replace the lens with another one, follow the procedure given below.
N
ote
• Only the lenses supplied as standard accessories with the VIVID 910 can be used. (They are not compatible with oth-
er VIVID models since they have been calibrated with the VIVID 910 only.)
• Since the VIVID 910 is a precision instrument, do not leave it without lens when replacing the lens. Replace the lens
as soon as possible.
• If you replace the lens before saving the image data, the image data will be lost. Before replacing the lens, all the
necessary operations (e.g. saving data) must be completed.
• Do not touch the lens surface with fingers.
• If dust or dirt has collected on the lens surface, remove it using a camera air blower etc.
• Dust or dirt on the lens mount surface may produce incorrect measurement data. In this case, wipe the lens mount
surface with a clean, dry cloth and then blow it with a camera air blower etc.
• Dust and dirt collected on the lens mount surface on the body side must also be removed with a camera air blower etc.
For display of the status information, refer to page 51.
[Replacing Procedure]
1
Remove the lens hood.
M
emo
The lens hood can be fixed with its bayonet mount.
2
Turn the lens fixing ring counter-clockwise to
loose it completely.
M
emo
The lens fixing ring can be fixed with its screw mount.
R
ef.
Vivid910-hard-E.book Page 27 Thursday, September 25, 2003 6:49 PM
Содержание Vivid 910
Страница 7: ...5 Chapter 1 Before Using the Instrument Vivid910 hard E book Page 5 Thursday September 25 2003 6 49 PM...
Страница 13: ...11 Chapter 2 Preparation Vivid910 hard E book Page 11 Thursday September 25 2003 6 49 PM...
Страница 26: ...24 Notes When Using the Instrument with Tripod Vivid910 hard E book Page 24 Thursday September 25 2003 6 49 PM...
Страница 27: ...25 Chapter 3 Operation Vivid910 hard E book Page 25 Thursday September 25 2003 6 49 PM...
Страница 54: ...52 Displaying the Status Information Vivid910 hard E book Page 52 Thursday September 25 2003 6 49 PM...
Страница 55: ...53 Chapter 4 Appendix Vivid910 hard E book Page 53 Thursday September 25 2003 6 49 PM...
Страница 63: ...Vivid910 hard E hyou fm Page 3 Thursday September 25 2003 6 52 PM...
Страница 64: ...2001 2006 KONICA MINOLTA SENSING INC AGKCKK 7 Printed in Japan 9222 1888 40 8KXKF JCTF J QW HO...