SKR 4000 DIRECT DIGITIZER User Manual
82
..................................................................................................................................................KONICA
MINOLTA, INC.
450
430-470
FM, ±5kHz deviation, 1kHz sine, 28V/m
710
704-787
Pulse modulation 217Hz, 9V/m
745
780
810
800-960
Pulse modulation 18Hz, 28V/m
870
930
1720
1700-1990
Pulse modulation 217Hz, 28V/m
1845
1970
2450
2400-2570
Pulse modulation 217Hz, 28V/m
5240
5100-5800
Pulse modulation 217Hz, 9V/m
5500
5785
Input AC power Port
Phenomenon
Basic EMC
standard
Immunity test levels
Professional healthcare facility
environment
Electrical
fast
transients/burst
IEC 61000-4-4
±2 kV
100kHz repetition frequency
Surges
Line-to-line
IEC 61000-4-5
±0.5 kV, ±1 kV
Surges
Line-to-ground
IEC 61000-4-5
±0.5 kV, ±1 kV, ±2 kV
Conducted
disturbances
induced
by RF fields
IEC 61000-4-6
3V, 0.15MHz-80MHz
6V in ISM bands between 0.15MHz and
80MHz
80%AM at 1kHz
Voltage dips
IEC 61000-4-11
0% UT; 0.5 cycle
At 0º, 45º,
90º, 135º, 180º, 225º, 270º and
315º
0% UT; 1 cycle
and
70% UT; 25/30 cycles
Single phase: at 0º
Voltage interruptions
IEC 61000-4-11
0% UT; 250/300 cycles
Input DC power Port
Phenomenon
Basic EMC
standard
Immunity test levels
Professional healthcare facility environment