Operation
KONICA MINOLTA,
INC..………….……………………………………………………………………………………………………
….…………….………..49
4.1.1.2. Work flow
Click “Prep”
Panel preparation, Wait until the warning message change from “Exposure
Prohibit” to “Exposure Enable”
Shoot X ray at any time. However the longer time you wait. The worse
image would be. So please shoot once x ray generator is ready.
After shooting, Click “Acquire”
Wait for image uploaded
4.1.1.3. Timing Setting
To set a clear scenario for programming, see diagram below for details
Workstation(Inner2)
HVG
Tube
1.
Workstation receives “prep” request, send command “Clear” to panel.
2.
Panel receives “clear” from workstation, starts clearing leakage of panel. Meanwhile, panel send a
message to workstation “Exposure Prohibited”.
3.
Panel finishes “Clear” and send a message to workstation “Exposure Enable”.
4.
Workstation shows “Exposure Enable” on the IDetector’s message bar to tell user shoot X ray now.
5.
User triggers x ray generator to initialize and do anode rotation to prepare for X ray shooting.
6.
X ray generator finishes preparation for X ray shooting and reminds user to shoot.