35
Symptom
Displayed message
Possible cause
Correction
CM-
36dG
CM-
36d
The specified operation
cannot be executed
because motor drive is in
progress.
The operation cannot
be executed because
the motor that performs
switching of the
measurement area and
SCI/SCE is currently
operating.
Wait for motor operation to
end before executing the
operation.
There was an operating
failure of the A/D
conversion device inside
the instrument.
The A/D conversion device
installed in the instrument
has failed.
Turn the power OFF and
then turn it ON again. If
the message is displayed
again, contact a KONICA
MINOLTA-authorized
service facility.
There was an operating
failure of light output
circuit charging inside the
instrument.
Charging of the light
output circuit is not
completed. The light
output circuit is not
operating correctly.
There was an operating
failure of the light
output circuit inside the
instrument.
The light output circuit
installed in the instrument
is not operating correctly.
There was an operating
failure of a motor inside the
instrument.
The motor that performs
switching of the
measurement area,
SCI/SCE, and gloss
measurement is not
operating correctly.
There was an operating
failure of memory inside
the instrument.
The memory installed
in the instrument is not
operating correctly.
The camera used to capture
viewfinder images is not
operating.
The camera installed in the
instrument is not operating
correctly.