The gloss calibration plate is used when performing gloss calibration.
Install the target mask to use for measurement in advance.
From the software in advance, set the same measurement area, specular component (SCI/SCE), and UV output
that will be used for measurement.
Procedure
1 Pull the sample holder toward you and hold it so it is
opened.
2 As shown in the figure, press on the sample holder so that
the sample holder fits into the indentation on the reverse
side of the gloss calibration plate.
Installing the Gloss Calibration Plate
²
Precautions for Use of the Gloss Calibration Plate
Memo
• Gloss calibration data is used when performing gloss calibration using the gloss calibration plate. Gloss calibration data
is set in the instrument at the time when it is purchased.
• The gloss calibration plate may become discolored if left exposed to light. Therefore, make sure to close the cap
when the plate is not in use in order to prevent the plate from being exposed to external light.
• Be careful that the gloss calibration plate does not become scratched and does not contact fingerprints or other
dirt.
• If the gloss calibration plate becomes dirty, use a blower to blow off the dust, then wipe the dirt off using the
standard accessory cleaning cloth.
• If the dirt on the calibration plate does not come off easily, wipe using a soft cloth dampened with ethanol.
• If a part other than the calibration plate becomes dirty, gently wipe the dirt off with a soft cloth dampened with
water or soapy water.
• If the gloss calibration plate is scratched or the dirt cannot be removed, replace it. After the gloss calibration
plate was replaced, set the gloss calibration data to the data for the new gloss calibration plate.