111
De
scr
ipti
on
Error message
Cause (Description)
Corrective Action
7
OVER
Luminance of measuring
object is higher than avail-
able measurement range.
• Use ND filter and redo measure-
ment.
• If symptom does not improve,
please contact the nearest Konica
Minolta-authorized service facility.
8
OFFSET ERROR
Zero calibration has not
been correctly performed.
• Reset the power and redo meas-
urement.
• If symptom does not improve,
please contact the nearest Konica
Minolta-authorized service facility.
9
EXCESSIVE LUMI-
NANCE VARIATION
Luminance change for
measuring object is large.
• Place this instrument on stable
surface and redo measurement.
• If symptom does not improve,
please contact the nearest Konica
Minolta-authorized service facility.
10
VIEWING-ANGLE
SELECTOR ERROR
Measurement was per-
formed when measurement
angle selector was in wrong
position, or measurement
angle selector was changed
during measurement.
• Switch measurement angle selec-
tor and check that measurement
angle appears in screen (------ is
not displayed), and then redo
measurement. In addition, do not
operate measurement angle selec-
tor during measurement.
• If symptom does not improve,
please contact the nearest Konica
Minolta-authorized service facility.
11
ROM ERROR
Data stored in ROM is
broken.
• Do not switch OFF (O side) while
storing data or changing setting, or
when message “PLEASE WAIT...”
appears.
• If symptom does not improve,
please contact the nearest Konica
Minolta-authorized service facility.
CS-200_E101-118.indd 111
12.7.17 4:52:16 PM
Содержание CHROMA METER CS-200
Страница 21: ...19 Installing CS 200_E001 026 indd 19 12 7 17 4 43 25 PM...
Страница 29: ...27 Setting CS 200_E027 060 indd 27 12 7 17 4 46 00 PM...
Страница 57: ...55 Setting BACK LIGHT CS 200_E027 060 indd 55 12 7 17 4 46 28 PM...
Страница 62: ...60 CS 200_E027 060 indd 60 12 7 17 4 46 32 PM...
Страница 63: ...61 Measurement Preparation CS 200_E061 86 indd 61 12 7 17 4 47 50 PM...
Страница 88: ...86 CS 200_E061 86 indd 86 12 7 17 4 48 17 PM...
Страница 89: ...87 Measurement CS 200_E087 96 indd 87 12 7 17 4 49 12 PM...
Страница 98: ...96 CS 200_E087 96 indd 96 12 7 17 4 49 24 PM...
Страница 99: ...97 Communication CS 200_E97 100 indd 97 12 7 17 4 51 08 PM...
Страница 102: ...100 CS 200_E97 100 indd 100 12 7 17 4 51 09 PM...
Страница 103: ...101 Description CS 200_E101 118 indd 101 12 7 17 4 52 11 PM...
Страница 124: ...9222 1892 A1 2005 KONICA MINOLTA INC BIIBPK Printed in Japan...