66
M
ea
su
re
m
en
t
Pr
ep
ara
tio
n
(1) Through measurement
7.
Press either
or
key to select
[USER CAL] and then
ENTER
key.
< USER CAL DATA> screen
appears.
8.
Use close-up lens, select measuring
angle, adjust finder diopter and focus.
For more details on each operation, see p.88.
Use light source of which luminance and
chromaticity are known on object.
9.
Press
COLOR
key in SHIFT mode,
and select color space.
Select same color space as calibration value
to be entered in procedures from
12
to
13
.
10.
Press measurement button
to start measurement.
Now starts measurement.
This measurement is performed with
correction coefficient (CH00) based upon
KONICA MINOLTA's calibration standard.
Measurement value appears on
<CURRENT DATA>
part of LCD screen
after measurement is completed.
After measurement, do not switch OFF
until measurement value appears.
Doing so may break stored data.
continuation from a p.64, 65.
9
16
11, 15
13
7, 12
9
U
S
E
R C
A
L
<
CH SETTING
>
CH01:
<
DEFAULT
>
USER CAL
CH ID NAME
CH COPY
CH RESET
BREAK:[ESC]
[ ]
=
[ 2° ]
<USER CAL DATA>
CH01:<DEFAULT>
Lv ------
cd/m
2
x ------
y ------
SELECT MEH DATA
OR MEASURE.
NEXT : [ENTER]
[ SFT]
=
[ 2° ]
<
ANGL
>
1
°
VIEW
<
SPD
>
AUTO
<
SYNC
>
NO SYNC
<
LENS
>
STANDARD
<
MEM
>
AUTO NUM
<HOLD DATA>
Lv ------
x ------
y ------
<USER CAL DATA>
CH01:<DEFAULT>
Lv ------
cd/m
2
x ------
y ------
SELECT MEH DATA
OR MEASURE.
NEXT : [ENTER]
[ SFT]
=
[ 2° ]
<
ANGL
>
1
°
VIEW
<
SPD
>
AUTO
<
SYNC
>
NO SYNC
<
LENS
>
STANDARD
<
MEM
>
AUTO NUM
<CURRENT DATA>
Lv 90.30
x 0.4210
y 0.4128
<USER CAL DATA>
CH01:<DEFAULT>
Lv ------
cd/m
2
x ------
y ------
SELECT MEH DATA
OR MEASURE.
NEXT : [ENTER]
[ ]
=
[ 2° ]
<
ANGL
>
1
°
VIEW
<
SPD
>
AUTO
<
SYNC
>
NO SYNC
<
LENS
>
STANDARD
<
MEM
>
AUTO NUM
<HOLD DATA>
Lv ------
x ------
y ------
CS-200_E061-86.indd 66
13.6.28 6:14:17 PM
Содержание CHROMA METER CS-200
Страница 21: ...19 Installing CS 200_E001 026 indd 19 12 7 17 4 43 25 PM...
Страница 29: ...27 Setting CS 200_E027 060 indd 27 12 7 17 4 46 00 PM...
Страница 57: ...55 Setting BACK LIGHT CS 200_E027 060 indd 55 12 7 17 4 46 28 PM...
Страница 62: ...60 CS 200_E027 060 indd 60 12 7 17 4 46 32 PM...
Страница 63: ...61 Measurement Preparation CS 200_E061 86 indd 61 12 7 17 4 47 50 PM...
Страница 88: ...86 CS 200_E061 86 indd 86 12 7 17 4 48 17 PM...
Страница 89: ...87 Measurement CS 200_E087 96 indd 87 12 7 17 4 49 12 PM...
Страница 98: ...96 CS 200_E087 96 indd 96 12 7 17 4 49 24 PM...
Страница 99: ...97 Communication CS 200_E97 100 indd 97 12 7 17 4 51 08 PM...
Страница 102: ...100 CS 200_E97 100 indd 100 12 7 17 4 51 09 PM...
Страница 103: ...101 Description CS 200_E101 118 indd 101 12 7 17 4 52 11 PM...
Страница 124: ...9222 1892 A1 2005 KONICA MINOLTA INC BIIBPK Printed in Japan...