51
S
e
tt
in
g
5.
Press either
or
key
to switch update method of
measurement value directory.
If
key is pressed, mode switches in order of [AUTO
NUM]
→
[AUTOSAVE]
→
[MAN NUM]
→
[AUTO
NUM]. If kept pressed, this switches continuously.
In case of
key, mode switches in reversing
order. Keep pressing to switch continuously.
With [AUTO NUM] setting,
memory channel number is replaced automatically
after storing data by pressing
MEMORY
key. For example, if M005 is displayed,
measurement value is to store in M005 after
MEMORY
key is pressed and number
of memory channel displayed in
<MEMORY DATA>
is replaced to M006.
If [AUTOSAVE] is set,
measurement value will be saved automatically after
measurement and memory channel number changes accordingly. For example, if M005
is displayed, measurement value is to store in M005 after measurement is performed
and number of memory channel displayed in
<MEMORY DATA>
is replaced to M006.
In case of [MAN NUM],
measurement value is to store memory channel of which number
is shown then. For example, if M005 is displayed, measurement value is to store in M005
and nunber of memory channel displayed in
<MEMORY DATA>
remaines M005.
With [AUTO NUM] or [AUTOSAVE],
and M100 has been used up, data goes back to M000 again.
When data storage is to perform on memory channel with another
measurement value, a warning message appears.
Warning message “OK TO OVERWRITE?” appears. If OK, press
ENTER
key, and if not,
ESC
key.
Setting is available not to display such message. See p.50.
Press
ESC
key to stop.
6.
Press
ENTER
key.
Inversion cursor returns to
[MEMORY MODE
]. Now change has been set.
7.
Press
ESC
key to return to
measurement screen.
remains even after switching OFF (O).
M
E
M
O
R
Y
M
O
D
E
CH00:
<
DEFAULT
>
Lv 20.80
cd/m
2
x 0.4476
y 0.4477
<
MODE
>
SINGLE
[ ]
=
[ 2° ]
<
ANGL
>
1
°
VIEW
<
SPD
>
AUTO
<
SYNC
>
200.00Hz
<
LENS
>
STANDARD
<
MEM
>
AUTOSAVE
<
MEMORY DATA
>
M
OOO
:(NO DATA)
Lv ------
x ------
y ------
<
MENU
>
2/4
OBJECTIVE LENS STANDARD
DISPLAY DIGITS 4 FIGURES
MEMORY MODE
AUTO NUM
DATA PROTECT ON
BREAK : [ESC]
[ ]
=
[ 2° ]
<
MENU
>
2/4
OBJECTIVE LENS STANDARD
DISPLAY DIGITS 4 FIGURES
MEMORY MODE
AUTOSAVE
DATA PROTECT ON
BREAK : [ESC]
[ ]
=
[ 2° ]
CS-200_E027-060.indd 51
12.7.17 4:46:25 PM
Содержание CHROMA METER CS-200
Страница 21: ...19 Installing CS 200_E001 026 indd 19 12 7 17 4 43 25 PM...
Страница 29: ...27 Setting CS 200_E027 060 indd 27 12 7 17 4 46 00 PM...
Страница 57: ...55 Setting BACK LIGHT CS 200_E027 060 indd 55 12 7 17 4 46 28 PM...
Страница 62: ...60 CS 200_E027 060 indd 60 12 7 17 4 46 32 PM...
Страница 63: ...61 Measurement Preparation CS 200_E061 86 indd 61 12 7 17 4 47 50 PM...
Страница 88: ...86 CS 200_E061 86 indd 86 12 7 17 4 48 17 PM...
Страница 89: ...87 Measurement CS 200_E087 96 indd 87 12 7 17 4 49 12 PM...
Страница 98: ...96 CS 200_E087 96 indd 96 12 7 17 4 49 24 PM...
Страница 99: ...97 Communication CS 200_E97 100 indd 97 12 7 17 4 51 08 PM...
Страница 102: ...100 CS 200_E97 100 indd 100 12 7 17 4 51 09 PM...
Страница 103: ...101 Description CS 200_E101 118 indd 101 12 7 17 4 52 11 PM...
Страница 124: ...9222 1892 A1 2005 KONICA MINOLTA INC BIIBPK Printed in Japan...