Selecting Lens Type
Use optional accessory close-up lens for small area measurement. See
instruction manual for close-up lens for the placement of one. If close-up lens is
to use, measurement value is required for calibration of lens transmission factor.
Since calibration value varies depending on lens type, lens type has to be set in
this instrument in advance. Erroneous setting causes incorrect measurement.
Below table shows lens to set and setting for this instrument.
Lens to Set
No lens
Close-Up lens
No.107
Close-Up lens
No.122
Arbitrary lens like optional accessory ND
filter (1/10) CS-A6, ND Filter (1/100) CS-A7,
Combination of close-up lens and ND filter
Setting
STANDARD
No.107
No.122
USER1
∗
USER2
∗
USER3
∗
∗
Setting at the shipment from factory : STANDARD
∗
Use standard accessory data management software CS-S10w Standard or optional
CS-S10w Professional to avail more lens type like USER1, USER2 and USER3, and
register calibration value for lens other than close-up lens No.107 and No.122 set in
USER1, USER2 or USER3. See instruction manual for data management software for
details.
Operation Procedure
1.
Press
ESC
key when menu or target
value setting menu appears.
Measurement screen
appears on LCD screen.
2.
Make sure that [ ] appears on left
bottom indicating this instrument
has not been set for SHIFT mode.
If
[SFT]
appears instead, it means that
instrument is to operate in SHIFT mode.
Press
SHIFT
key to cancel SHIFT mode.
2
3
1, 7
6
4, 5
O
B
JE
C
T
IV
E
LE
N
S
CH00:
<
DEFAULT
>
Lv 20.80
cd/m
2
x 0.4476
y 0.4477
<
MODE
>
SINGLE
[ ]
=
[ 2° ]
<
ANGL
>
1
°
VIEW
<
SPD
>
AUTO
<
SYNC
>
200.00Hz
<
LENS
>
STANDARD
<
MEM
>
AUTO NUM
<
MEMORY DATA
>
M
OOO
:(NO DATA)
Lv ------
x ------
y ------
CS-200_E027-060.indd 40
12.7.17 4:46:15 PM
Содержание CHROMA METER CS-200
Страница 21: ...19 Installing CS 200_E001 026 indd 19 12 7 17 4 43 25 PM...
Страница 29: ...27 Setting CS 200_E027 060 indd 27 12 7 17 4 46 00 PM...
Страница 57: ...55 Setting BACK LIGHT CS 200_E027 060 indd 55 12 7 17 4 46 28 PM...
Страница 62: ...60 CS 200_E027 060 indd 60 12 7 17 4 46 32 PM...
Страница 63: ...61 Measurement Preparation CS 200_E061 86 indd 61 12 7 17 4 47 50 PM...
Страница 88: ...86 CS 200_E061 86 indd 86 12 7 17 4 48 17 PM...
Страница 89: ...87 Measurement CS 200_E087 96 indd 87 12 7 17 4 49 12 PM...
Страница 98: ...96 CS 200_E087 96 indd 96 12 7 17 4 49 24 PM...
Страница 99: ...97 Communication CS 200_E97 100 indd 97 12 7 17 4 51 08 PM...
Страница 102: ...100 CS 200_E97 100 indd 100 12 7 17 4 51 09 PM...
Страница 103: ...101 Description CS 200_E101 118 indd 101 12 7 17 4 52 11 PM...
Страница 124: ...9222 1892 A1 2005 KONICA MINOLTA INC BIIBPK Printed in Japan...