100
401027/B
About dimensional surveying and alignment
The EM 712 Multibeam echo sounder is a precision instrument for scientific
measurements.
In order to obtain precision data that are both detailed and correct, it is necessary to align
the transducer, measure the location and offset of each sensor in relation to the vessel’s
coordinate system, and calibrate the complete EM 712 prior to use. Modest accuracy
requirements apply when your EM 712 system is only used to investigate objects in
the water column.
The quality assurance tasks required for EM 712 include:
• Aligning the transducer during installation
• Dimensional surveying
• Calibration
The alignment and dimensional surveying must be done during the EM 712 installation
with the vessel in dry dock. The first calibration is normally done at sea during the Sea
Acceptance Test. This calibration may not be complete, and must then be repeated later.
The calibration is then repeated at regular interfaces, and prior to each survey.
Dimensional surveying
Determining the relative positions and orientations of the sensors and the transducer with
high accuracy is important. This can only be met using a survey company/personnel
with good experience in maritime dimensional surveying.
The dimensional surveying tasks required for EM 712 include:
1
Establishing the vessel coordinate system
2
Defining the location of the origin in the coordinate system
3
Setting out the required coordinate reference points throughout the vessel.
4
Defining the vessel’s centre line, and if required — identify this line with physical
markings
5
Measuring the necessary position and angles for all relevant sensors in the
coordinate system.
6
Measuring the location and orientation of the EM 712 transducer in the coordinate
system.
All results from the dimensional survey measurements must be summarized in a report
by the consultants doing the work.
The information provided by the dimensional survey is entered into the EM 712 software
as installation parameters.
Kongsberg EM 712 Installation manual
Содержание EM 712
Страница 1: ...401027 B June 2018 Kongsberg Maritime AS Kongsberg EM 712 Multibeam echo sounder Installation manual ...
Страница 68: ...68 401027 B Processing Unit Topics CBMF board dip switch setting page 69 Kongsberg EM 712 Installation manual ...
Страница 71: ...401027 B 71 216148 EM 712 Transducer TX1 dimensions Drawing file ...
Страница 72: ...72 401027 B Kongsberg EM 712 Installation manual ...
Страница 73: ...401027 B 73 221048 EM 712 Transducer TX2 dimensions Drawing file ...
Страница 74: ...74 401027 B Kongsberg EM 712 Installation manual ...
Страница 75: ...401027 B 75 219621 EM 712 Transducer RX1 dimensions Drawing file ...
Страница 76: ...76 401027 B Kongsberg EM 712 Installation manual ...
Страница 77: ...401027 B 77 216146 EM 712 Transducer RX2 dimensions Drawing file ...
Страница 78: ...78 401027 B Kongsberg EM 712 Installation manual ...
Страница 79: ...401027 B 79 223137 EM 712 Transducer mounting frame 0 5 degrees Drawing file ...
Страница 80: ...80 401027 B Kongsberg EM 712 Installation manual ...
Страница 81: ...401027 B 81 223139 EM 712 Transducer mounting frame 1 degree Drawing file ...
Страница 82: ...82 401027 B Kongsberg EM 712 Installation manual ...
Страница 83: ...401027 B 83 223273 EM 712 Transducer mounting frame 2 degrees Drawing file ...
Страница 84: ...84 401027 B Kongsberg EM 712 Installation manual ...
Страница 85: ...401027 B 85 317812 EM 712 Casing w mounting frame 0 5 degrees Drawing file ...
Страница 86: ...86 401027 B 320320 EM 712 Casing w mounting frame 1 degree Kongsberg EM 712 Installation manual ...
Страница 87: ...401027 B 87 375817 EM 712 Combined casing w mounting frame 1 degree Drawing file ...
Страница 88: ...88 401027 B 331369 EM 712 Casing w mounting frame 2 degrees Kongsberg EM 712 Installation manual ...
Страница 89: ...401027 B 89 396402 EM 712 Transmitter Unit dimensions Drawing file ...
Страница 90: ...90 401027 B 212984 EM 712 Transmitter Unit mounting bracket Kongsberg EM 712 Installation manual ...
Страница 91: ...401027 B 91 396428 EM 712 Receiver Unit dimensions Drawing file ...
Страница 92: ...92 401027 B 385422 Processing Unit dimensions Kongsberg EM 712 Installation manual ...
Страница 93: ...401027 B 93 378828 Hydrographic Work Station dimensions Drawing file ...
Страница 94: ...94 401027 B Kongsberg EM 712 Installation manual ...
Страница 95: ...401027 B 95 371591 Rack installation kit dimenisons Drawing file ...
Страница 96: ...96 401027 B 370275 Remote Control Unit K REM dimensions Kongsberg EM 712 Installation manual ...
Страница 97: ...401027 B 97 Drawing file ...
Страница 98: ...98 401027 B 373962 Remote Control Unit K REM wiring diagram Kongsberg EM 712 Installation manual ...
Страница 112: ...112 401027 B RX array identification RX2 TX Unit identification 3 2 RX 3 4 2 RX 4 Kongsberg EM 712 Installation manual ...
Страница 163: ...401027 B 163 Certificates IEC 60945 IACS E10 Technical specifications ...
Страница 177: ...401027 B 177 7 Observe the handling rules for transducers Equipment handling ...
Страница 188: ... 2018 Kongsberg Maritime ...