go on until the settled measure times. The equipment will check the standard
deviation in each steps.
Min. Meas Times: 2-99
Max. Meas Times: 2-99
Deviation/20m: the average standard deviation in 20m
Eg. Measure Range= 60m, Standard Deviation/20m = 0.007m
Standard = Deviation x 60m = 0.0021m
In 60m, the max. Standard deviation value is 0.0021m
Median: Middle value for measurement in several times, for example n=3 (2...99)
If n is odd, for example the value is 2,5,6, the median is 5
If n is even, for example the value is 2,5,6,7, the median is 5.5
3.6 Function
Press [SHIFT] + [FNC] under any measure interface to enter the function setting page.
If a function is unavailable when calling-out, it means this function is not suitable for
current application.
3.6.1 Test
The Test function afford a solution for users who want to measure in several times without
saving data. It can also test the equipment or improve the distance measurement. The Test
function is independent to current job.
3.6.2 View
In View page, it can show the last measured result in current job.