15.2 Circular Vial
Inspection
No adjustment is necessary if the bubble of the circular vial is in the center after inspection
and adjustment of the plate vial.
Adjustment
If the bubble of the circular vial is not in the center bring the bubble to the center by using
the adjusting pin or hexagon wrench to adjust the bubble adjusting screw. First loosen the
screw opposite to the offset side and then tighten the other adjusting screw on the offset
side, bringing the bubble to the center. After the bubble stays in the center each of the
three adjustment screws should be tightened in a uniform manner.
15.3 Compensator
After leveling the instrument accurately, the tilt angle should be close to 0, otherwise it
exist zero error of tilt sensor, which will affect the measurement result.
Inspection
1. Leveling instrument accurately.
2. Open the XY in the E bubble page, details refer to 3.3
3. Read compensation tilt angle value X1 and Y1 after the display stable
4.Rotate telescope 180°, read the compensation tilt angle value X2 and Y2 after the
display stable