TF.16 IFU StarDevice.003.10-EN 2021-01-12
prepared by Pulavskyi A., Kolyesnik O.
34
Pulse frequency
The pulse frequency of the manipulator is set by the doctor from 4 to 12 shocks
per second. The doctor can change the frequency value during work at his own
discretion. It is forbidden to change the settings during the correction session
(when the manipulator knocks).
Pulse force
The intensity or force of the correction can be set at the range from 4.54 kg / cm2
(Value 10) to 15.87 kg / cm2 (value 35).
There are the following recommended settings of force of a shock for different areas
of the spine:
For the cervical spine: 10-15
For the thoracic spine: 15-25
For the lumbar spine: 25-35
The doctor chooses the settings for other joints / soft tissues at his own discretion.
5.9 SUBMENU
You need to press the central button «OK» to enter submenu.
The submenu allows you to change the number of shocks in the therapy
session, to change the load index, to reset the shock counters and to calibrate
the manipulator.
You can change the parameters of submenu by using the red carriage. The current settings are marked in
green. The carriage is moving up / down by using the F + and F- buttons respectively. Moves to the right
/ left are performed by using the buttons FR + and FR-, respectively. To apply the parameters, you need
to move the carriage to the next line and to press the central button «OK».