6
.
Installation
(1)
Installation place
①
This equipment is designed for indoor use.
Never install the equipment in a place where it is likely to be wetted by water or rain. The
equipment may fail. Install the equipment in a place where sufficient ventilation is provided
and changes in humidity are minimal.
②
Install the equipment in a place free of vibration and impact.
③
Do not use the equipment under direct sunlight or at high temperature/humidity.
④
Install the equipment in a place free of dust.
⑤
Install the equipment in a place free of corrosive gases.
⑥
Install the equipment in a place free of a strong electric/magnetic field.
⑦
Install the equipment in a place where ambient temperature is 15 to 35
℃
.
The use of the equipment in a condition deviating from the conditions described in the
specifications is a cause of failure.
When installing the mass flow unit, noise generating sources in the
vicinity of installation, environment filled with moisture and dust and very
hot, corrosive gas ambient must be avoided. Such conditions are causes of
serious failure.
(2)
Installation method
Install the equipment using the threaded holes (M4) on the bottom of the body.
For installation hole dimensions, refer to 19. External View on page 33.
(3)
Installation posture
The basic installation posture is horizontal. If the equipment is to be installed vertically,
warm up the equipment fully and conduct the zero point adjustment.
Determine the direction of gas inlet and outlet as indicated by the arrows on the ID plate.
CAUTION
Содержание 8700 MC
Страница 1: ...MJ400306A1 1 Thermal Mass Flow Meter Controller MODEL 8700 SERIES INSTRUCTION MANUAL...
Страница 33: ...MJ400306A1 33 19 External View 1 Integrated indicator type...
Страница 34: ...MJ400306A1 34...
Страница 35: ...MJ400306A1 35 2 Separate indicator type Main unit...
Страница 36: ...MJ400306A1 36...
Страница 37: ...MJ400306A1 37 Indicator URL http www kofloc co jp...