K441XX Automatic Non-Woods Metal Noack Evaporative Apparatus  

Operation and Instruction Manual 

 

K441XX-Manual 

- 12 - 

otherwise, even an incomplete analysis running 
longer than 5 minutes will be recorded.   

 
3.  The status bar will describe the progress of the 

analysis. In the bottom right hand side of the 
screen the graph will display the temperature 
data. 

 
4.  An alarm will sound along with a full screen 

notice three minutes before the end of the 
analysis. See Figure 12 below. Touch the 
screen to shut off the alarm and close the full 
screen notice. The status bar will display the 
countdown to the end of the analysis. 
 

 

 

Figure 12: End Analysis Screen 

 
5.  As the analysis is approaching the end, prepare 

3 cm deep water bath at ambient temperature.  
This will be used to cool down the crucible; 
therefore it should be large enough to hold a 
sufficient amount of water. Sufficient cool down 
of the crucible will take approximately 30 
minutes so accommodate a water bath where 
the crucible can safely rest without interfering 
with regular laboratory activities.  (A laboratory 
sink would be sufficient so long as no 
interference occurs). 

 
6.  When the analysis time reaches 60 minutes 

(the countdown reaches 0), a new alarm will 
sound.  Touch the screen to stop the alarm, 
then CAREFULLY remove the crucible and 
place it in the above mentioned water bath.  Let 
the crucible rest in the bath for 30 minutes. 

 

NOTE: WHILE REMOVING THE CRUCIBLE 
AND PLACING IT IN THE WATER BATH 
MAKE SURE THE SAMPLE DOES NOT 
SPLASH THE CRUCIBLE’S COVER. 
 

7.  After 30 minutes, carefully dry the outside of the 

crucible then, unscrew the cover. 

NOTE:

 This step should be performed without 

tilting the crucible.  

THE SAMPLE SHOULD 

NOT SPLASH THE CRUCIBLE’S COVER. 

 

8.  Reweigh the crucible with the remaining (un-

evaporated) part of the sample to the nearest 
0.01g. 

 
9. In 

the 

Results Browser

 enter the final weight.  

The analyzer will calculate the final result. 
 

6.4  Analysis of the Results 

 

Subtract the empty crucible’s weight from the end 

result in order to maintain the weight of the 
remaining sample.  The result of the following 
formulas will provide the value of Noack 
Evaporation Loss (Test Method D5800) 

 

Method A:

 

                                            (B-A) – (C-A) 
 

Evaporation Loss =   

――――――――

 

 

 

 

                   B-A 

 
 Where: 
 

 

A= Empty Crucible weight 

 

 

B= Crucible plus Sample weight 
C= Crucible plus Sample after 1   
      hour of heating 

 

Method B:

 

 
 [(M

1

-M

2

)/M

1

] x100 

 

 

 Where: 

 

M

1

= sample weight before the test 

 

 

M

2

= sample weight after the test 

 

The final result is given to the nearest 0.1% 
on both methods 

 

NOTE:

 The analyzer is able to perform all 

calculations on the above mentioned formulas, the 
operator is only required to: 

 

  Note and introduce the empty 

crucible’s weight 

  Note and introduce the sample’s 

weight before the test 

  Note and introduce the crucible’s 

plus sample’s weight after the test 

 

The analyzer automatically gives the result 
correlation between the two procedures as 
prescribed in the ASTM D5800 method: 
 
Result on B= 1.030 x Result on A 
Result on A= 0.970 x Result on B 

Содержание K44100

Страница 1: ...mia New York 11716 1796 USA Toll Free 1 800 878 9070 US only Tel 1 631 589 3800 Fax 1 631 589 3815 http www koehlerinstrument com e mail info koehlerinstrument com Petroleum Testing Analysis Instrumen...

Страница 2: ...1...

Страница 3: ...Packing List 5 3 2 Unpacking 5 3 3 Installation 5 3 4 Additional Accessories Required for Running Tests 5 3 5 Optional Accessories Available 5 4 Descriptions 6 4 1 Piping Diagram 6 4 2 Instrument Cont...

Страница 4: ...s manual should also be used in conjunction with applicable published laboratory procedures Information on these procedures is given in section 1 2 1 1 Koehler s Commitment to Our Customers Providing...

Страница 5: ...to access the analyzer STAY ALERT Do not operate this unit while under the influence of medication alcohol or drugs Equipment Modifications and Replacement Parts Any modification or alteration of thi...

Страница 6: ...shipping container and equipment Do not return goods to Koehler without written authorization 3 3 Installation Equipment Placement Place the instrument on a firm surface in a secure area 1 Connect th...

Страница 7: ...6 4 Descriptions 4 1 Piping Diagram Figure 1 Piping Diagram...

Страница 8: ...K441XX Automatic Non Woods Metal Noack Evaporative Apparatus Operation and Instruction Manual K441XX Manual 7 4 2 Instrument Controls and Connections Figure 2 Top Front Left Side View...

Страница 9: ...K441XX Automatic Non Woods Metal Noack Evaporative Apparatus Operation and Instruction Manual K441XX Manual 8 Figure 3 Top Left Back Side View Figure 4 Analyzer Complete with Assembled Glassware Set...

Страница 10: ...bration date Figure 7 Calibration Screen 4 On the left side of the screen a graph shows the calibration curve along with the selected probe s value in both C and volts 5 On the right side of the scree...

Страница 11: ...emperature Control via PT100 probe in the Heating Block Bath Method B Temperature Control via PT100 probe in the Crucible 4 Select the set point temperature for the test based on the selected method S...

Страница 12: ...egulate the aspiration vacuum to 2 mbar by slowly closing the bleed valve halfway and regulating the vacuum regulator on the analyzer s side in order to reach 20 0 2mm on the inclined manometer 2mbar...

Страница 13: ...for 30 minutes NOTE WHILE REMOVING THE CRUCIBLE AND PLACING IT IN THE WATER BATH MAKE SURE THE SAMPLE DOES NOT SPLASH THE CRUCIBLE S COVER 7 After 30 minutes carefully dry the outside of the crucible...

Страница 14: ...e operator to select the sample o Press the Load Analysis Button in between the two arrows to view the analysis Figure 13 Results Browser Main Screen If the final weight has not been introduced a ques...

Страница 15: ...hler Representative as each printer has different characteristics and specific drivers Touching the Show Details button causes the analysis to reload and the data to be displayed second by second on t...

Страница 16: ...test is run clean the Woullf bottles and pipes Each time a test is run clean the heating jacket Periodically clean the vacuum filter 7 2 Replacement Parts Some instrument parts may need to be replaced...

Страница 17: ...rize the return of the product to the factory The sole responsibility of Koehler Instrument Company and the purchaser s exclusive remedy for any claim arising out of the purchase of any product is the...

Страница 18: ...K441XX Automatic Non Woods Metal Noack Evaporative Apparatus Operation and Instruction Manual K441XX Manual 17 Notes...

Страница 19: ...K441XX Automatic Non Woods Metal Noack Evaporative Apparatus Operation and Instruction Manual K441XX Manual 18 Notes...

Страница 20: ...K441XX Automatic Non Woods Metal Noack Evaporative Apparatus Operation and Instruction Manual K441XX Manual 19 Notes...

Отзывы: