K441XX Automatic Non-Woods Metal Noack Evaporative Apparatus  

Operation and Instruction Manual 

 

K441XX-Manual 

- 10 - 

6 Operation 

 

6.1  Summary of Test Method 

 

A quantity of 65 grams of sample is heated in a 

crucible at 250°C with a constant flow of air drawn 
through it for 60 minutes. At the end of the test, the 
sample is cooled and reweighed.  The loss in mass 
of the oil is then determined. 

 

 

6.2  Analysis Setup Procedure 

 

WARNING:

 DO NOT INSERT THE CRUCIBLE 

INTO THE BATH UNTIL THE SOFTWARE 
PROMPTS THE USER AT THE START OF 
OPERATION.  
 
1. Press the 

Start Analysis

 button on the main 

menu screen (Figure 5) to open the analysis 
startup panel. See Figure 8 below: 

 

 

 

Figure 8: Analysis Startup Screen 

 
2. Touch the 

Analysis Name

 field to identify the 

sample.  An on-screen keyboard will pop up 
(See Figure 9 below).  The analysis name 
entered will automatically be listed as the test 
title in the results browser. This will allow the 
operator to easily recognize the analysis when 
using the results browser for review. 

 

 

 

Figure 9: On-Screen Keyboard Pop-Up 

NOTE:

 If the analysis name is left blank the 

analysis will not start. 
 

3.  Select the method by touching the 

Method

 

field.  Method A is shown in Figure 8 above.  

 

Method A:

 Temperature Control via PT100 

probe in the Heating Block (Bath)  

Method B:

 Temperature Control via PT100 

probe in the Crucible. 

 
4.  Select the set-point temperature for the test 

based on the selected method.  Switching the 
method will automatically change the 
temperature to that prescribed by the method. 
 

5.  Weigh the empty crucible without the cover to 

the nearest 0.01g and enter the measured 
weight.   
 

6.  Pour 65g ± 0.1g of sample into the crucible and 

enter the measured weight to the nearest 0.01g 
into the dedicated fields. 
 

7.  When all of the above parameters are entered, 

start the analysis by pressing the 

START

 

button. 

 
WARNING:

 DO NOT INSERT THE CRUCIBLE 

INTO THE BATH UNTIL THE SOFTWARE 
PROMPTS THE USER AT THE START OF 
OPERATION.  

 

8.  According to the method and temperature 

selected, the bath will heat to the prescribed 
set-point.   

 
NOTE: 

The initial heating phase will show the 

bath’s temperature even if method B was 
selected. During this phase the graph will not 
display data and the aspiration will remain idle.   

 
9.  While the bath is heating, ensure the glassware 

is correctly set-up as shown on Page Figure 1, 
Page 6.  Verify that the manometer is level and 
fill with water as needed so that the inclined 
scale reads zero.  

Make sure the bleed valve 

is fully open.

 

 

10. Once the bath has reached the set temperature 

an audible alarm will sound and the Crucible 
Prompt will appear on the screen. See Figure 
10: 

 

 

Содержание K44100

Страница 1: ...mia New York 11716 1796 USA Toll Free 1 800 878 9070 US only Tel 1 631 589 3800 Fax 1 631 589 3815 http www koehlerinstrument com e mail info koehlerinstrument com Petroleum Testing Analysis Instrumen...

Страница 2: ...1...

Страница 3: ...Packing List 5 3 2 Unpacking 5 3 3 Installation 5 3 4 Additional Accessories Required for Running Tests 5 3 5 Optional Accessories Available 5 4 Descriptions 6 4 1 Piping Diagram 6 4 2 Instrument Cont...

Страница 4: ...s manual should also be used in conjunction with applicable published laboratory procedures Information on these procedures is given in section 1 2 1 1 Koehler s Commitment to Our Customers Providing...

Страница 5: ...to access the analyzer STAY ALERT Do not operate this unit while under the influence of medication alcohol or drugs Equipment Modifications and Replacement Parts Any modification or alteration of thi...

Страница 6: ...shipping container and equipment Do not return goods to Koehler without written authorization 3 3 Installation Equipment Placement Place the instrument on a firm surface in a secure area 1 Connect th...

Страница 7: ...6 4 Descriptions 4 1 Piping Diagram Figure 1 Piping Diagram...

Страница 8: ...K441XX Automatic Non Woods Metal Noack Evaporative Apparatus Operation and Instruction Manual K441XX Manual 7 4 2 Instrument Controls and Connections Figure 2 Top Front Left Side View...

Страница 9: ...K441XX Automatic Non Woods Metal Noack Evaporative Apparatus Operation and Instruction Manual K441XX Manual 8 Figure 3 Top Left Back Side View Figure 4 Analyzer Complete with Assembled Glassware Set...

Страница 10: ...bration date Figure 7 Calibration Screen 4 On the left side of the screen a graph shows the calibration curve along with the selected probe s value in both C and volts 5 On the right side of the scree...

Страница 11: ...emperature Control via PT100 probe in the Heating Block Bath Method B Temperature Control via PT100 probe in the Crucible 4 Select the set point temperature for the test based on the selected method S...

Страница 12: ...egulate the aspiration vacuum to 2 mbar by slowly closing the bleed valve halfway and regulating the vacuum regulator on the analyzer s side in order to reach 20 0 2mm on the inclined manometer 2mbar...

Страница 13: ...for 30 minutes NOTE WHILE REMOVING THE CRUCIBLE AND PLACING IT IN THE WATER BATH MAKE SURE THE SAMPLE DOES NOT SPLASH THE CRUCIBLE S COVER 7 After 30 minutes carefully dry the outside of the crucible...

Страница 14: ...e operator to select the sample o Press the Load Analysis Button in between the two arrows to view the analysis Figure 13 Results Browser Main Screen If the final weight has not been introduced a ques...

Страница 15: ...hler Representative as each printer has different characteristics and specific drivers Touching the Show Details button causes the analysis to reload and the data to be displayed second by second on t...

Страница 16: ...test is run clean the Woullf bottles and pipes Each time a test is run clean the heating jacket Periodically clean the vacuum filter 7 2 Replacement Parts Some instrument parts may need to be replaced...

Страница 17: ...rize the return of the product to the factory The sole responsibility of Koehler Instrument Company and the purchaser s exclusive remedy for any claim arising out of the purchase of any product is the...

Страница 18: ...K441XX Automatic Non Woods Metal Noack Evaporative Apparatus Operation and Instruction Manual K441XX Manual 17 Notes...

Страница 19: ...K441XX Automatic Non Woods Metal Noack Evaporative Apparatus Operation and Instruction Manual K441XX Manual 18 Notes...

Страница 20: ...K441XX Automatic Non Woods Metal Noack Evaporative Apparatus Operation and Instruction Manual K441XX Manual 19 Notes...

Отзывы: