- 9 -
5 Instrument
Calibration
1. From the main menu screen (See Figure 5
below) select Diagnostic Maintenance.
Figure 5: Diagnostic Maintenance Panel
2. In
the
Diagnostic Maintenance Panel
(See
Figure 6 below) select
Calibration.
This will
open the calibration panel.
Figure 6: Diagnostic Maintenance Panel
3. At the top of the Calibration Screen (See Figure
7) three buttons will allow the user to choose
which probe/ sensor will be displayed. The
Calibrated in
label directly below these buttons
shows the selected probe’s last calibration date.
Figure 7: Calibration Screen
4. On the left side of the screen, a graph shows
the calibration curve along with the selected
probe’s value in both °C and volts.
5. On the right side of the screen, the
Offset
field
allows for fine tuning of the calibration.
NOTE:
If an offset has been entered, the
original calibration curves will be shown as a
green line. The black line will move according
to the offset value entered in real time.
6. The
Perform Calibration
button starts the
guided calibration procedure: to perform a
calibration of a temperature probe it is
necessary to use a RTD/PT100 Simulator. The
recommended RTD/PT100 simulator is the
KLA-PT100-CAL Calibrated Decade Box with
KLA-DB-KIT Connectors and Cables.