Operation and Maintenance
3-1
2000 March Rev. B 8754020
Operation and Maintenance
System Power Up
1. Set the imager Power Switch to ON. (The Local Panel Power light will turn on.)
The imager performs a series of self-diagnostic tests and begins a warmup
cycle. (The time required for warmup varies based on how long the imager
has been turned off. The maximum warmup time at normal room temperature
is about 25 minutes.)
After the diagnostic tests and warmup cycle are complete, the Local Panel
displays SO1 Ready and the Ready light turns ON.
!
Caution
On occasion the imager may have to perform file maintenance on its hard disk
after power on. When this occurs, the message P154 Disk Maint displays on
the Local Panel. The process may require several minutes. To avoid longer
delays, NEVER turn off power during disk maintenance. As long as the P154
Disk Maint message is displayed, the system is operating correctly.
2. You can acquire images during the warmup cycle, but the system will not print
until the Ready light turns on.
Power Switch
(Rear Cabinet)
Local
Panel
Power
Light
Ready
Light
Figure 3-1. Power Switch and Local Panel