D-4
A-61167 April 2005
Calibrating the flatbed
Calibration optimizes the optical system in your scanner in order to achieve
the best overall quality of scanned images. Frequent calibration is not needed
or recommended.
NOTES: You must calibrate the ADF before you can calibrate the flatbed.
The screens shown in this section are for the TWAIN driver. Your
screens may be different.
1. Allow the lamps to warm up for three minutes.
2. Click on
Calibrate
in the Imaging tab.
The Image Chain Calibration dialog box appears.
3. Click
Calibrate
.
A message appears.
4. Place the calibration target in the scanner ADF.
Use the 29.7 cm (11.7 in.) square calibration target (CAT. No. 131 7304)
that is included with your scanner.
5. Click
OK
.