6-7
Tables of Exposure Times
The following exposure timetables were established with a unit equipped with
a standard cone with a focus-to-skin distance of 20 cm (8 inches).
70kV – 7mA
Cone 20 cm (8 in.)
60kV – 7mA
Cone 20 cm (8 in.)
Maxillary
Child
Adult
Large
Maxillary
Child
Adult
Large
Incisor / Bicuspid
0.119
0.180
0.227
Incisor / Bicuspid
0.238
0.360
0.454
Premolar
0.152
0.230
0.290
Premolar
0.304
0.460
0.580
Molar
0.178
0.270
0.340
Molar
0.356
0.540
0.680
Mandibular
Child
Adult
Large
Mandibular
Child
Adult
Large
Incisor / Bicuspid
0.099
0.150
0.189
Incisor / Bicuspid
0.198
0.300
0.378
Premolar
0.106
0.160
0.202
Premolar
0.211
0.320
0.403
Molar
0.119
0.180
0.227
Molar
0.238
0.360
0.454
Bitewing
Child
Adult
Large
Bitewing
Child
Adult
Large
Anterior
0.099
0.150
0.189
Anterior
0.198
0.300
0.378
Posterior
0.119
0.180
0.227
Posterior
0.238
0.360
0.454
Occlusal
Child
Adult
Large
Occlusal
Child
Adult
Large
0.205
0.310
0.391
0.409
0.620
0.781
Table 7: Exposure times in seconds for class D-Speed films
Setting
Correction
Coefficient
S 01
- 92 %
0.08
S 02
- 90 %
0.10
S 03
- 88 %
0.125
S 04
- 84 %
0.16
S 05
- 80 %
0.20
S 06
- 75 %
0.25
S 07
- 68 %
0.31
S 08
- 60 %
0.40
S 09
- 50 %
0.50
S 10
- 37 %
0.63
S 11
- 21 %
0.79
S 12
0 %
1.00
S 13
+ 26 %
1.26
S 14
+ 59 %
1.59
S 15
+ 100 %
2.00
S 16
+ 152%
2.52
Table 8: Exposure time corrections depending on selected setting
Содержание 2200
Страница 1: ...KODAK 2200 Intraoral X ray System User s Manual ...
Страница 10: ...SM731_K2200_01_en 06 2009 1 6 ...
Страница 24: ...SM731_K2200_01_en 06 2009 3 6 ...
Страница 26: ...SM731_K2200_01_en 06 2009 4 2 ...
Страница 30: ......
Страница 36: ...SM731_K2200_01_en 06 2009 6 6 Figure 15 Heating and cooling curves of the KODAK 2200 system tube head ...
Страница 41: ...06 2009 SM731_K2200_01_en 6 11 ...