06/2009 SM731_K2200_01_en
5-3
Troubleshooting
Problem
Cause
Solution
Nothing lights up
Unit is disconnected.
Connect the unit.
Fuse F1 is burned out or defective.
Replace the fuse.
Circuit breaker is off.
Turn on the circuit breaker.
Control unit doesn’t light
up
Control unit is disconnected.
Connect the control unit.
Fuse F1 is burned out or defective.
Replace the fuse.
Control unit is defective.
Call a qualified service technician.
No x-ray emission
Generator is cooling.
Wait for the I 01 message and the red
LED on the control unit to disappear.
Radiology control key is defective.
Call a qualified service technician.
X-ray emission works, but
exposure is too light or
completely white.
Wrong exposure pre-set program is
being used.
Change the pre-set program (refer to the
User Mode section).
Generator is positioned incorrectly.
Adjust the position of the generator.
Exposure time is too short.
Increase the exposure time.
Development time is too short.
Increase the development time (refer to
the development instructions).
Developer is too cold.
Heat the developer.
Developer is too old or diluted.
Replace with fresh developer.
Pre-set mode is incorrectly selected.
Verify your exposure settings (refer to
the exposure procedure).
Receptor is facing the wrong way.
Reposition the receptor.
Unit was incorrectly installed.
Call a qualified service technician.
X-ray emission works, but
exposure is too dark.
Wrong exposure pre-set program is
being used.
Change the pre-set program (refer to the
User Mode section).
Exposure time is too long.
Decrease the exposure time.
Development time is too long.
Decrease the development time (refer to
the development instructions).
Developer is too hot.
Cool the developer.
Developer is too concentrated.
Adjust the concentration or replace the
developer.
Pre-set mode is incorrectly selected.
Verify your exposure settings (refer to
the exposure procedure).
Table 4: Troubleshooting
Содержание 2200
Страница 1: ...KODAK 2200 Intraoral X ray System User s Manual ...
Страница 10: ...SM731_K2200_01_en 06 2009 1 6 ...
Страница 24: ...SM731_K2200_01_en 06 2009 3 6 ...
Страница 26: ...SM731_K2200_01_en 06 2009 4 2 ...
Страница 30: ......
Страница 36: ...SM731_K2200_01_en 06 2009 6 6 Figure 15 Heating and cooling curves of the KODAK 2200 system tube head ...
Страница 41: ...06 2009 SM731_K2200_01_en 6 11 ...