06/2009 SM731_K2200_01_en
5-1
5. Care and Maintenance
General Maintenance
To make sure that the system functions correctly, you must have it serviced
annually by an authorized technician. In addition, every three months inspect
the equipment and make sure of the following:
Generator
•
The certification label is legible.
•
There are no oil leaks.
Mechanical support
•
The wall framework is securely attached to the wall.
•
All the labels are legible.
•
The scissor arm is stable in all positions.
Control unit and electrical installation
•
The symbols are legible.
•
The control unit cable and the power supply cable are in good
condition.
•
The ground is correctly installed.
•
The radiology control key returns to its initial position after use.
Functioning
•
The audible signal is audible and the x-ray emission light is visible
when you make an exposure (manual mode, 70kV, 0.1 sec.).
•
The message “E 01”, which means Operator Error, is displayed when
you make an exposure (manual mode, 70kV, 1.0 sec.) and release the
control button before the exposure time has elapsed.
Timer self-test
•
Turn on the system to activate the self-test.
•
The test starts with a simultaneous test of the display and alarm
lights.
•
The unit proceeds to the systems test. At the end of this test,
indicated by a short beep, the firmware version and the total number
of exposures (divided by 10) made by the machine since installation
is displayed.
•
If the test is not successful, an Error code is displayed on the display.
Important
If the result of any of these checks is
not satisfactory, discontinue using the
equipment and contact an authorized
technician.
Содержание 2200
Страница 1: ...KODAK 2200 Intraoral X ray System User s Manual ...
Страница 10: ...SM731_K2200_01_en 06 2009 1 6 ...
Страница 24: ...SM731_K2200_01_en 06 2009 3 6 ...
Страница 26: ...SM731_K2200_01_en 06 2009 4 2 ...
Страница 30: ......
Страница 36: ...SM731_K2200_01_en 06 2009 6 6 Figure 15 Heating and cooling curves of the KODAK 2200 system tube head ...
Страница 41: ...06 2009 SM731_K2200_01_en 6 11 ...