6. Acquire the image.
•
Press the x-ray exposure button on the control timer unit (or on the
remote exposure switch). The x-ray emission indicator lights up and
an audible signal is emitted.
•
Keep pressing until the x-ray emission light goes out and the audible
signal stops.
CAUTION:
If you stop pressing the control key before the exposure ends, a
manipulator alarm is activated. It indicates that the x-ray emission was
interrupted prematurely and that there is a risk of underexposure.
7. Read the emitted dose.
Quickly press the selection knob. The “mGy” indicator lights up and the
dose in mGy is displayed. Section 6 provides a table with emitted dosage
based on exposure times (see Table 12).
Processing
When using conventional film, process the film according to manufacturer’s
instructions. Develop the film under safelight conditions in an automatic
processor or manually.
If you use an automatic processor, refer to the processor’s manual. Be sure
to maintain the mechanically and keep the solutions replenished.
If you develop film manually, follow precisely the manufacturer’s
recommendations for solution preparation, development time, and solution
temperature. Any deviation from the manufacturer’s recommendations (such
as a solution that is too concentrated or diluted, too hot or cold, or if film is
processed for the wrong amount of time), will adversely affect the quality of
the final radiograph.
Setting Modes
1. Select the mode to be set
by pressing and holding the selection
button for at least 3 seconds, as many time as necessary to get the
desired LED lighted.
2.
Press and hold simultaneously
for at least 3 seconds the selection
button and the knob until the display blinks. The current setting is
displayed.
3.
Modify the setting
by rotating the knob until the new setting is
displayed (refer to the setting table as indicated in Tables 9 to 11).
4. Validate your choice
by pressing and holding the knob until the
message “copy” is displayed or
abort
by pressing shortly the knob.
Содержание 2200
Страница 1: ...KODAK 2200 Intraoral X ray System User s Manual ...
Страница 10: ...SM731_K2200_01_en 06 2009 1 6 ...
Страница 24: ...SM731_K2200_01_en 06 2009 3 6 ...
Страница 26: ...SM731_K2200_01_en 06 2009 4 2 ...
Страница 30: ......
Страница 36: ...SM731_K2200_01_en 06 2009 6 6 Figure 15 Heating and cooling curves of the KODAK 2200 system tube head ...
Страница 41: ...06 2009 SM731_K2200_01_en 6 11 ...