SM731_K2200_01_en
06/2009
6-2
Rated high voltage and maximum corresponding current
•
70 kV, 7 mA
Current/voltage combinations for a maximum output power of:
•
490 W, 70 kV / 7 mA
Rated power for exposure time of 0.1 sec.
•
490 W
Rate of use
•
At 70 kV, 7 mA and 0.1 sec. and at the maximum tank temperature: one
exposure every 8 sec.
Minimum value of the current/time product in the range of
conformity
•
0.07 mAs at 7 Ma
Selection of parameters
•
70 kV / 7 mA and 60 kV / 7 mA
Area of conformity to IEC standard 60601-2-7
•
Reproducibility of the emitted radiation: conform
•
Linearity of the emitted radiation: conform
•
Precision in radiography: conform
Measurement conditions
•
kV: Indirect measurement using a kV peakmeter
•
mAs: Direct measurement in the circuit using a mAs-meter
•
Exposure time: Indirect measurement on the kV signal at 75% of the
peak value
Type of protection against
electric shock
Class 1 equipment
Degree of protection against
electric shock
Type B
Protection against harmful
ingress of water
Ordinary equipment
Operation mode
Continuous operation with
intermittent loading
Flammable anesthetics
Not suitable for use in presence of
flammable anesthetics or a
mixture of flammable anesthetics
with air or oxygen or nitrous oxide
Содержание 2200
Страница 1: ...KODAK 2200 Intraoral X ray System User s Manual ...
Страница 10: ...SM731_K2200_01_en 06 2009 1 6 ...
Страница 24: ...SM731_K2200_01_en 06 2009 3 6 ...
Страница 26: ...SM731_K2200_01_en 06 2009 4 2 ...
Страница 30: ......
Страница 36: ...SM731_K2200_01_en 06 2009 6 6 Figure 15 Heating and cooling curves of the KODAK 2200 system tube head ...
Страница 41: ...06 2009 SM731_K2200_01_en 6 11 ...