9. Parameters – Description and Programming
9.1 Measurement Configuration
P00: - cba Application/Engineering Units
Programming of this parameter will result in loading the factory default with the
corresponding engineering units.
a
Operating (measurement) mode
0
Liquid level measurement
b
Engineering units
(according to “c”)
Attention: mind the sequence!
When programming this parameter the
right value “a” will be blinking first.
Metric
US
0
m ft
1
cm inch
c
Calculation system
0
Metric
1
US
FACTORY DEFAULT: 000
P01: - ba Measurement Mode – Bargraph
Parameter value „a” will determine the basic measurement value that will be displayed and
proportional with the current output. Depending on the value of “a” process values as listed in
the 3d column can also be displayed by pressing NEXT . For return to the display of the
basic value the ENTER
E
key should be pressed.
a
Measurement
Mode
Display
symbol
Displayed values
Attention: mind the
sequence!
When programming this
parameter the
right value “a” will be
blinking first.
0
Distance
DIST
Distance
1
Level
LEV
Level, Distance
2
Level in
percentage
LEV%
Level%, Level, Distance
3
Volume
VOL
Volume, Level, Distance
4
Volume in
percentage
VOL%
Volume%, Volume, Level,
Distance
5
Flow
FLOW
Flow, TOT1, TOT2, Level,
Distance
Parameter value “b” will determine that the height of the Bargraph will be proportional to the
current output or to the Echo strength.
b
Bargraph indication
0
Echo strength
1
Current output
FACTORY DEFAULT: 11