The NUS-4 will be delivered with the following Factory Default:
Current output, display and bargraph: LEVEL
Current output and bargraph proportional to the level
4 mA: assigned to the minimum level 0 %
20 mA: assigned to the maximum level 100 %
Error indication by the current output: hold last value
Damping: 60 sec
The display module supports two separately accessible programming modes
representing 2-layers of programming complexity, depending on user choice.
PLUG-IN DISPLAY MODULE
QUICKSET
FULL
PARAMETER
ACCESS
8.3.1
Display Module NUS-400P
Symbols used on the LCD
DIST
– Distance
(measuring) mode
LEV
– Level (measuring)
mode
VOL
– Volume
(measuring) mode
FLOW
– Open channel
(flow metering) mode
PROG
- Programming
mode (device under
programming)
RELAY
– ‘C2’ circuit of
the relay is closed
T1
- TOT1 volume flow
totaliser (resetable
aggregate)
T2
- TOT2 volume flow
totaliser (aggregate)
FAIL
- Measurement /
device error
- Level changing
direction
Bargraph assigned to the current
output or echo strength
Symbols used on the frame
M
– Metric system
US
– US calculation
system
LEDs lit
VALID
– presence of
valid echo