NUS-4
NUS-4 K05/0216
Seite 17
8.3.6
Full Parameter Access
Full Parameter Access is the highest programming level to access all features:
Example:
Setting the measurement
Programming of the output
Optimization of measurement
Selection of 11 pre-programmed tank shapes for volume measurement
Selection of 20 pre-programmed measurement channels for the flow
measurement in open channels
Keys
Function
ENTER
E
+ NEXT
(press for 3 seconds)
Enter or exit Full Parameter Access programming mode.
In this programming mode, the display will indicate:
yy
Parameter Adress (
P01
,
P02
…
P99
)
xxxx Parameter value (dcba)
Bar graph
Note: Measuring is going on during programming in accordance with
the old parameter set. New parameter set will be valid after returning
to the Measurement to the Programming Mode.
Steps and indications of the Full Parameter Access programming mode
pressing
Keys
while
Parameter Address is blinking
while
Parameter Value is blinking
ENTER
E
Go to the Parameter Value
Save the modification of the
Parameter Value and return to the
Parameter Address
NEXT + UP
Cancel all modifications of the actual
programming phase. Pressing for 3
sec is required while CANCEL will be
displayed for warning
Neglect the modification of the
Parameter Value.
and return to the Parameter Address
without saving the modifications
NEXT + DOWN Reset entire device to Factory
Default.
Since this action will reset all
parameters, “LOAD” will appear on
the display:
- to confirm, press
- to escape, press any other key
- Exception: clearing TOT 1 (See at
P77)
Display default of the Parameter
Values
(it can be saved by pressing ENTER
E
)
NEXT
Move blinking (changeability) of the digit to the left
UP / DOWN
Modify the blinking digit (increase, decrease) or scroll up/down
yy:xxxx