![Kobold NGM-HART Скачать руководство пользователя страница 30](http://html1.mh-extra.com/html/kobold/ngm-hart/ngm-hart_operating-instructions-manual_1982600030.webp)
Configuration Set NGM-HART
page 30
NGM-HART K05/0922
5.4.16
Signal Discussion 2
Level Coaxial Probe
• Stable reference reflection at the beginning
• Negative level reflection at 168mm
• No end of probe reflection as energy is completely reflected at water surface
Dead band parameter at 30 mm.
20 mA parameter at 50 mm.
4 mA parameter at 230 mm.