Configuration Set NGM-HART
page 16
NGM-HART K05/0922
5.4.2
Amplitude threshold
BASIC CONFIGURATION
• Read out actual amplitude threshold, by clicking on I19
Field G19 indicates the actual amplitude threshold. Dynamic noise or ringing can be blocked if it is
within the amplitude threshold band. The level reflection should be always 1/3 bigger than the
width of the amplitude threshold band.
• Change actual amplitude threshold in field G18 and click on I18 “set amplitude
threshold”.
Verify changes by clicking again on I19 “get amplitude threshold”