Configuration Set NGM-HART
NGM-HART K05/0922
page 29
5.4.15
Signal Discussion 1
Empty Coaxial Probe
• Nice reference reflection at the beginning
• Perfect coupling into the coaxial probe
• Positive end of probe reflection which corresponds to the physical end of probe
Dead band parameter at 30 mm.
20 mA parameter at 50 mm.
4 mA parameter at 230 mm.
Amplitude threshold at 1000.