Parameter table
Output 1/2 – Part volume
Sub-
parameter
level 1
Sub-
parameter
level 2
Sub-
parameter
level 3
Description
Value range /
value list
Standard
value LPM
Standard
value
GPM
Disabled
Output deactivated
disabled
Pulse output
Pulse unit
List
selection
Sets the Pulse
Volume for Pulse
Volume
ml, L, m3,
galUS, galUK,
User
L
galUs
Pulse
volume
Value input
Sets the value for the
pulse volume
0-999
1
Pulse width
Sets the pulse width 1-20.000
1 ms
Control input
(only Out1)
Control function for
counter reset
OFF, memory
reset
Off
KofiCom
(only OUT1)
Factory calibration mode on output 1
IO-Link
(only OUT1)
This mode activates the IO-Link function on output 1
10.7 User service
The user service provides the user with a reset function and password setting.
Together with the activation of a user password, therefore, the menu access for
the user on the part of a master user can be blocked.
10.7.1
User service / change password
In the factory setting the user password is set to "00000", the user functions are
thus freely accessible. If the user password is changed to other than "00000", the
password prompt becomes active the next time the user menu is entered.
If the set password is no longer known, a master password can be requested from
KOBOLD.
10.7.2
User service / factory setting
By activating this function, the user can reset the device to the factory settings. Any
user settings will then be lost and the device will be back in delivery condition.