MIM-
page 20
MIM 01/0419
10.4.1 Flow
10.4.1.1 Unit
The displayed unit for the flow measurement can be selected from various
predefined standard units. It is also possible to define a user-defined unit ("user"),
here the "user unit" must be in LPM (liters / min.) be programmed:
e.g. Unit User = 100 LPM, if Q = 500 LPM then the display shows 5 users.
10.4.1.2 Separation
The parameter Separation sets the flow rate below which the measured value is
set to "0". If this function is active, the flow value "0" is shown in blue colour in the
display.
10.4.1.3 Simulation
mode
See section 10.4.5
10.4.2 Volume
10.4.2.1 Counter
type
Absolute:
Regardless of the flow direction, the calculated partial volume is added to the
counters.
Bidirectional:
Depending on the flow direction, the calculated partial volume is added or
subtracted to the counters. If the measured flow value is negative, the volume value
goes down from measurement to measurement (possibly into the negative range).
10.4.2.2
Unit of the total volume counter
The parameter "Unit" determines the volume unit of the total volume counter. The
listed volume units are available. When changing the volume unit, the current
counter reading is converted to the new volume unit.