Kobold HND-P210 Скачать руководство пользователя страница 9

 

 

HND-P210 

HND-P210 K08/0422 

 

page 9 

 

SEt 
CLOC

 

 

Set Clock: Setting Of Real Time Clock

 

CLOC

 

HH:MM

 

Clock

: Setting of time hours:minutes

 

dAtE

 

TT.MM

 

Date

: day.month

 

YEAr

 

YYYY

 

Year

 

 

Note: If the logger memory contains data already, the 
menues/parameters marked with (*) cannot be invoked! If these 
should be altered the logger memory has to be cleared before! 
(key 6, see chapter 

8.6 Operation Of Logger

)

 

 

8.4.1 

Different Kinds of Measuring: “rAtE-Slo, -P.dEt, -FASt” 

Three different kinds of measuring pressure are supported. Two of them (P.dEt 
and FASt) are working with high measuring frequency of more than 1000 
measurings per second. 
 

8.4.1.1 

rAtE-Slo:   

Standard Measuring 

Measuring rate 4Hz, averaging and filter functions are active. 
Application: Measuring of slowly changing or static pressures, eg. measuring of 
leakproofness, atmospheric pressure... 
Highest accuracy, high noise immunity (EMI and unstable measuring signals), 
low power consumption. 
 

8.4.1.2 

rAtE-P.dEt: 

Peak detection 

Measuring rate >1000Hz, the value is displayed unfiltered. 
Application with logger function: Measuring of short pressure peaks or fast 
changing pressures with a resolution of <1ms. The cyclic logger function records 
the arithmetic mean value, the highest and the lowest peak of the referring time 
interval. 
Attention: higher power consumption, measuring is sensitive to noise (EMI,..) 
 

8.4.1.3 

rAtE-FASt: 

Fast filtered measuring 

Measuring rate >1000Hz, the value is filtered slightly (higher noise immunity than 
P.dEt., small peaks will be filtered out), apart from that identical behaviour like 
P.dEt. 

8.4.2 

Averaging Function 

The  averaging  function  concerns  the  display  values  (LCD  and  interface).  It  is 
completely  independent  from  the  averaging  of  the  logger  function,  please  don’t 
mix them up!

 

The averaging integrates the measuring values during a selectable period of time 
and then calculates the average display value. It is independent from the selected 
kind  of  measuring  (slow,  fast,  peak  detect).  As  long  as  not  enough  values  are 
collected  (selected  averaging  time)  to  calculate  a  average  value,  the  upper 
display shows “----“, the lower display a ‘countdown‘.

 

During an active low-power-logging procedure the averring is always deactivated 
Function of min/max-value memory during averaging:

 

Содержание HND-P210

Страница 1: ...Operating Instructions for Hand held Pressure Measuring Devices for External Pressure Sensors Model HND P210...

Страница 2: ...r Level Display Unit m 11 8 6 Operation Of Logger 11 8 7 The Serial Interface 14 8 8 Pressure Connection to The Sensors 15 8 9 Error and System Messages 16 8 10 Calibration Services 16 9 Maintenance 1...

Страница 3: ...iced by persons familiar with these operating instructions and in accordance with local regulations applying to Health Safety and prevention of accidents When used in machines the measuring unit shoul...

Страница 4: ...eration With Power Supply Warning When using a power supply please note that operating voltage has to be 10 5 to 12 VDC Do not apply overvoltage Cheap 12 V power supplies often have excessive no load...

Страница 5: ...e the device temperature has adjusted to the ambient temperature before trying a new start up 4 If device is to be connected to other devices e g via serial interface the circuitry has to be designed...

Страница 6: ...Furthermore the kind of measuring is displayed Slo Fast P det p r t 2 1 and Corr if a Sensor with activated offset or scale correction is connected After changing the battery the clock setting menu is...

Страница 7: ...tment press button 3 for approx 5 seconds Auto Null will be displayed shortly The adjustment is done via the OFFSET value of the sensor referring configuration menu Note A zero point adjustment can on...

Страница 8: ...mbar Enter desired value at which the analogue output potential should be 1 V if Out dAC Set CAL Set Calibration Adjustment of Sensor OFFS Sensordep e g 5 00 5 00 mbar The offset of sensor will be dis...

Страница 9: ...Application with logger function Measuring of short pressure peaks or fast changing pressures with a resolution of 1ms The cyclic logger function records the arithmetic mean value the highest and the...

Страница 10: ...rface converter To get access to each device the base addresses of the devices have to be different For example choose 01 for the first 11 for the second device and so on See also chapter 8 7 The Seri...

Страница 11: ...tore 2 seconds The logger records 3 measurement results each time current or mean value depending on logger setting see below min peak and max peak of sensor Min and max peak are the minimum resp the...

Страница 12: ...etween the different values referring to the measurement also is done by pressing Changing the measurement is done by pressing the keys or 8 6 2 Func CYCL Automatic Recording With Selectable Logger Cy...

Страница 13: ...ording Do not stop the recording The selection can be made by key 2 and key 5 Quit key 6 enters the choice Note If you try to switch off the instrument in the cyclic recording operation You will be as...

Страница 14: ...me Conf P oFF x 12 Read ID number x 223 Set power off time Conf P oFF x 22 Read min alarm rail AL AL Lo x x 224 Logger Read data of CYCL Logger x 23 Read max alarm rail AL AL Hi x 225 Logger Read cycl...

Страница 15: ...e measuring range by re plugging the tube to pressure port A Please note that all values are displayed as positive values No minus sign will be shown Example for HND PS02 For tube connection B the mea...

Страница 16: ...Sensor defective Return to manufacturer for repair Err 2 Measured value below allowable range Check pressure not within sensor range measuring value to low Sensor defective Return to manufacturer for...

Страница 17: ...ensor Accuracy typ 0 1 F S at nominal temperature FASt and P dET 0 5 FS Measuring rate slow 4 meas sec ConF Rate slow fast 1000 meas sec ConF Rate FASt and P dEt Nominal temperature 25 C Sensor All se...

Страница 18: ...consumption Slow measuring rate 1 6mA Fast measuring rate 7 0mA Low Power Logger 0 3mA for cycle time 30s without interface communication active and no alarm horn sounding up to 0 8mA at cycle time 1s...

Страница 19: ...HND P210 HND P210 K08 0422 page 19 11 Order Codes Order no Housing design HND P210 1 x pressure sensor input with additional functions...

Страница 20: ...S06 0 2000 mbar abs HND PS07 0 7 00 bar abs 10 mbar max 10 bar abs HND PS08 0 400 0 mbar rel 0 1 mbar max 2 bar rel 0 70 C External threads G Edelstahl HND PS19 0 1000 mbar rel 1 mbar max 5 bar rel HN...

Страница 21: ...olated HND Z033 Adapter RS232 converter on USB interface HND Z034 Windows software for setting and data read and print out of instruments of the HND series with logger function BUS S20M Software for r...

Страница 22: ...ds noted below EN 61326 1 2013 Electrical equipment for measurement control and laboratory use EMC requirements Part 1 General requirements EN IEC 63000 2018 Technical documentation for the assessment...

Отзывы: