HND-P210
page 10
HND-P210 K08/0422
If averaging is activated and slow measuring is selected (rAtE-Slo), the min-
/max-value memory refers to the average display value.
If averaging is activated and fast measuring is selected (rAtE-FASt or P.dEt) ,
the min-/max-value memory refers to the internal measured values (fast peaks
can be detected).
8.4.3
Power Off Time
If there won’t be pressed any key and no interface communication takes place for
the time of the power off time setting (P.Off), the device will be switched off
automatically to save battery power.
If P.oFF = oFF then the automatic switch off is deactivated.
8.4.4
Address
Up to 10 devices of the HND-P- handheld-family can be connected to a serial
interface at once (depending on interface converter). To get access to each
device the base addresses of the devices have to be different. For example,
choose 01 for the first, 11 for the second device and so on. See also chapter
8.7
The Serial Interface
.
8.4.5
Alarm
There are three possible settings: Alarm off (
AL. oFF
), on with horn sound (
AL. on
),
on without sound (
AL. no.So
).
Following conditions will display an alarm, when the function is activated (on or
no.So):
Value is below lower (AL. Lo) or above upper alarm rail (AL.Hi).
Sensor error (Sens Erro)
Low battery (bAt)
Fe 7: System error (always with sound)
In case of an alarm and when polling the interface, the prio-flag is set in the
returned interface message.
8.4.6
Real Time Clock
The real time clock is used for the logger function: Recorded values are also
containing the point of time, when they were measured. Please check the settings
when necessary.
If the battery was replaced the refering menu ‚CLOC‘ will automatically be started.