  If averaging is activated and slow measuring is selected (rAtE-Slo), the min-

/max-value memory refers to the average display value. 

  If averaging is activated and fast measuring is selected (rAtE-FASt or P.dEt) , 

the min-/max-value memory refers to the internal measured values (fast peaks 
can be detected).

 

 

8.4.3 

Power Off Time 

If there won’t be pressed any key and no interface communication takes place for 
the  time  of  the  power  off  time  setting  (P.Off),  the  device  will  be  switched  off 
automatically to save battery power.  
If P.oFF = oFF then the automatic switch off is deactivated. 
 

8.4.4 

Address 

Up  to  10  devices  of  the  HND-P-  handheld-family  can  be  connected  to  a  serial 
interface  at  once  (depending  on  interface  converter).  To  get  access  to  each 
device  the  base  addresses  of  the  devices  have  to  be  different.  For  example, 
choose 01 for the first, 11 for the second device and so on. See also chapter 

8.7 

The Serial Interface

.

 

 

8.4.5 

Alarm 

There are three possible settings: Alarm off (

AL. oFF

), on with horn sound (

AL. on

), 

on without sound (

AL. no.So

). 

Following  conditions  will  display  an  alarm,  when  the  function  is  activated  (on  or 
no.So): 

  Value is below lower (AL. Lo) or above upper alarm rail (AL.Hi). 

  Sensor error (Sens Erro) 

  Low battery (bAt) 

  Fe 7: System error (always with sound) 

In  case  of  an  alarm  and  when  polling  the  interface,  the  prio-flag  is  set  in  the 
returned interface message.

 

 

8.4.6 

Real Time Clock 

The  real  time  clock  is  used  for  the  logger  function:  Recorded  values  are  also 
containing the point of time, when they were measured. Please check the settings 
when necessary. 
If the battery was replaced the refering menu ‚CLOC‘ will automatically be started.

 

 
 

Содержание HND-P210

Страница 1: ...Operating Instructions for Hand held Pressure Measuring Devices for External Pressure Sensors Model HND P210...

Страница 2: ...r Level Display Unit m 11 8 6 Operation Of Logger 11 8 7 The Serial Interface 14 8 8 Pressure Connection to The Sensors 15 8 9 Error and System Messages 16 8 10 Calibration Services 16 9 Maintenance 1...

Страница 3: ...iced by persons familiar with these operating instructions and in accordance with local regulations applying to Health Safety and prevention of accidents When used in machines the measuring unit shoul...

Страница 4: ...eration With Power Supply Warning When using a power supply please note that operating voltage has to be 10 5 to 12 VDC Do not apply overvoltage Cheap 12 V power supplies often have excessive no load...

Страница 5: ...e the device temperature has adjusted to the ambient temperature before trying a new start up 4 If device is to be connected to other devices e g via serial interface the circuitry has to be designed...

Страница 6: ...Furthermore the kind of measuring is displayed Slo Fast P det p r t 2 1 and Corr if a Sensor with activated offset or scale correction is connected After changing the battery the clock setting menu is...

Страница 7: ...tment press button 3 for approx 5 seconds Auto Null will be displayed shortly The adjustment is done via the OFFSET value of the sensor referring configuration menu Note A zero point adjustment can on...

Страница 8: ...mbar Enter desired value at which the analogue output potential should be 1 V if Out dAC Set CAL Set Calibration Adjustment of Sensor OFFS Sensordep e g 5 00 5 00 mbar The offset of sensor will be dis...

Страница 9: ...Application with logger function Measuring of short pressure peaks or fast changing pressures with a resolution of 1ms The cyclic logger function records the arithmetic mean value the highest and the...

Страница 10: ...rface converter To get access to each device the base addresses of the devices have to be different For example choose 01 for the first 11 for the second device and so on See also chapter 8 7 The Seri...

Страница 11: ...tore 2 seconds The logger records 3 measurement results each time current or mean value depending on logger setting see below min peak and max peak of sensor Min and max peak are the minimum resp the...

Страница 12: ...etween the different values referring to the measurement also is done by pressing Changing the measurement is done by pressing the keys or 8 6 2 Func CYCL Automatic Recording With Selectable Logger Cy...

Страница 13: ...ording Do not stop the recording The selection can be made by key 2 and key 5 Quit key 6 enters the choice Note If you try to switch off the instrument in the cyclic recording operation You will be as...

Страница 14: ...me Conf P oFF x 12 Read ID number x 223 Set power off time Conf P oFF x 22 Read min alarm rail AL AL Lo x x 224 Logger Read data of CYCL Logger x 23 Read max alarm rail AL AL Hi x 225 Logger Read cycl...

Страница 15: ...e measuring range by re plugging the tube to pressure port A Please note that all values are displayed as positive values No minus sign will be shown Example for HND PS02 For tube connection B the mea...

Страница 16: ...Sensor defective Return to manufacturer for repair Err 2 Measured value below allowable range Check pressure not within sensor range measuring value to low Sensor defective Return to manufacturer for...

Страница 17: ...ensor Accuracy typ 0 1 F S at nominal temperature FASt and P dET 0 5 FS Measuring rate slow 4 meas sec ConF Rate slow fast 1000 meas sec ConF Rate FASt and P dEt Nominal temperature 25 C Sensor All se...

Страница 18: ...consumption Slow measuring rate 1 6mA Fast measuring rate 7 0mA Low Power Logger 0 3mA for cycle time 30s without interface communication active and no alarm horn sounding up to 0 8mA at cycle time 1s...

Страница 19: ...HND P210 HND P210 K08 0422 page 19 11 Order Codes Order no Housing design HND P210 1 x pressure sensor input with additional functions...

Страница 20: ...S06 0 2000 mbar abs HND PS07 0 7 00 bar abs 10 mbar max 10 bar abs HND PS08 0 400 0 mbar rel 0 1 mbar max 2 bar rel 0 70 C External threads G Edelstahl HND PS19 0 1000 mbar rel 1 mbar max 5 bar rel HN...

Страница 21: ...olated HND Z033 Adapter RS232 converter on USB interface HND Z034 Windows software for setting and data read and print out of instruments of the HND series with logger function BUS S20M Software for r...

Страница 22: ...ds noted below EN 61326 1 2013 Electrical equipment for measurement control and laboratory use EMC requirements Part 1 General requirements EN IEC 63000 2018 Technical documentation for the assessment...

Отзывы: