Kobold HND-P210 Скачать руководство пользователя страница 12

HND-P210 

page 12 

 

HND-P210 K08/0422 

 

- time and date of the recording

 

After each recording „St. XX“ will be displayed for a short time. XX represents the 
number of the recording.

 

 

If logger memory contains recordings already: 

When „Store“ is pressed for 2 seconds, the choice for clearing the logger memory 
will be displayed:

 

 

Clear all 
recordings

 

Clear the last 
recording

 

 

Clear nothing 
(cancel menu)

 

 
The selection can be made by 

 (key 2) and 

 (key 5). "Quit" (key 6) enters the 

choice.

 

 

If the logger memory is full, the display will show: 

 

 

Viewing Recorded Measurings 

Within  the  „LoGG  Stor“  function  the  measurings  can  be  viewed  directly  in  the 
display not only by means of a computer (like at „Func CYCL“): press 2 seconds 
„Set“  (key  4):  The  first  menu  displayed  now  is  „rEAd  LoGG“  (read  logger  data). 
After  pressing 

  (key  3)  the  measurement  recorded  last  will  be  displayed, 

changing between the different values referring to the measurement also is done 
by pressing 

Changing the measurement is done by pressing the keys 

 or 

.

 

8.6.2 

„Func-CYCL“: Automatic Recording With Selectable Logger-

Cycle-Time 

The  Logger-Cycle-Time  is  settable  (p.r.t.  Configuration).  For  example  „CYCL“  = 
60: A measuring is recorded after each 60 seconds. 
When  the  slow  measurement  "rAtE-Slo"  is  chosen,  additionally  a  low  power 
function is available: „Lo.Po“. 
If „Lo.Po“ is on, the device only will take a measurement at the point of time of the 
recording. In between the recordings the measuring shuts down. This decreases 
the power consumption enormously and therefore is recommended e.g. for long 
time recordings where no mains adapter is available.

 

 
Max. number of measurings: 10000 
Cycle time: 1...3600 seconds (=1h), selectable in the configuration

 

 
A measuring contains: 



rAtE SLo:  

- current measuring value at the time of recording 

 

- min peak, max peak since the last recording 



rAtE FASt,P.dEt  

- arithmetic mean value since the last recording 

 

- min peak, max peak since the last recording

 

 

Содержание HND-P210

Страница 1: ...Operating Instructions for Hand held Pressure Measuring Devices for External Pressure Sensors Model HND P210...

Страница 2: ...r Level Display Unit m 11 8 6 Operation Of Logger 11 8 7 The Serial Interface 14 8 8 Pressure Connection to The Sensors 15 8 9 Error and System Messages 16 8 10 Calibration Services 16 9 Maintenance 1...

Страница 3: ...iced by persons familiar with these operating instructions and in accordance with local regulations applying to Health Safety and prevention of accidents When used in machines the measuring unit shoul...

Страница 4: ...eration With Power Supply Warning When using a power supply please note that operating voltage has to be 10 5 to 12 VDC Do not apply overvoltage Cheap 12 V power supplies often have excessive no load...

Страница 5: ...e the device temperature has adjusted to the ambient temperature before trying a new start up 4 If device is to be connected to other devices e g via serial interface the circuitry has to be designed...

Страница 6: ...Furthermore the kind of measuring is displayed Slo Fast P det p r t 2 1 and Corr if a Sensor with activated offset or scale correction is connected After changing the battery the clock setting menu is...

Страница 7: ...tment press button 3 for approx 5 seconds Auto Null will be displayed shortly The adjustment is done via the OFFSET value of the sensor referring configuration menu Note A zero point adjustment can on...

Страница 8: ...mbar Enter desired value at which the analogue output potential should be 1 V if Out dAC Set CAL Set Calibration Adjustment of Sensor OFFS Sensordep e g 5 00 5 00 mbar The offset of sensor will be dis...

Страница 9: ...Application with logger function Measuring of short pressure peaks or fast changing pressures with a resolution of 1ms The cyclic logger function records the arithmetic mean value the highest and the...

Страница 10: ...rface converter To get access to each device the base addresses of the devices have to be different For example choose 01 for the first 11 for the second device and so on See also chapter 8 7 The Seri...

Страница 11: ...tore 2 seconds The logger records 3 measurement results each time current or mean value depending on logger setting see below min peak and max peak of sensor Min and max peak are the minimum resp the...

Страница 12: ...etween the different values referring to the measurement also is done by pressing Changing the measurement is done by pressing the keys or 8 6 2 Func CYCL Automatic Recording With Selectable Logger Cy...

Страница 13: ...ording Do not stop the recording The selection can be made by key 2 and key 5 Quit key 6 enters the choice Note If you try to switch off the instrument in the cyclic recording operation You will be as...

Страница 14: ...me Conf P oFF x 12 Read ID number x 223 Set power off time Conf P oFF x 22 Read min alarm rail AL AL Lo x x 224 Logger Read data of CYCL Logger x 23 Read max alarm rail AL AL Hi x 225 Logger Read cycl...

Страница 15: ...e measuring range by re plugging the tube to pressure port A Please note that all values are displayed as positive values No minus sign will be shown Example for HND PS02 For tube connection B the mea...

Страница 16: ...Sensor defective Return to manufacturer for repair Err 2 Measured value below allowable range Check pressure not within sensor range measuring value to low Sensor defective Return to manufacturer for...

Страница 17: ...ensor Accuracy typ 0 1 F S at nominal temperature FASt and P dET 0 5 FS Measuring rate slow 4 meas sec ConF Rate slow fast 1000 meas sec ConF Rate FASt and P dEt Nominal temperature 25 C Sensor All se...

Страница 18: ...consumption Slow measuring rate 1 6mA Fast measuring rate 7 0mA Low Power Logger 0 3mA for cycle time 30s without interface communication active and no alarm horn sounding up to 0 8mA at cycle time 1s...

Страница 19: ...HND P210 HND P210 K08 0422 page 19 11 Order Codes Order no Housing design HND P210 1 x pressure sensor input with additional functions...

Страница 20: ...S06 0 2000 mbar abs HND PS07 0 7 00 bar abs 10 mbar max 10 bar abs HND PS08 0 400 0 mbar rel 0 1 mbar max 2 bar rel 0 70 C External threads G Edelstahl HND PS19 0 1000 mbar rel 1 mbar max 5 bar rel HN...

Страница 21: ...olated HND Z033 Adapter RS232 converter on USB interface HND Z034 Windows software for setting and data read and print out of instruments of the HND series with logger function BUS S20M Software for r...

Страница 22: ...ds noted below EN 61326 1 2013 Electrical equipment for measurement control and laboratory use EMC requirements Part 1 General requirements EN IEC 63000 2018 Technical documentation for the assessment...

Отзывы: