9

 

Conditions of use 

The VDI/VDE guidelines must be observed.  The meter is suitable for liquid products with sufficient flowability 
that are free of solids, do not bond and do not tend to settle and gases with linear flow behavior and an adequate 
inlet pressure. 
KDS-R: The minimum differential pressure between input and output side must amoun to 350 mbar. 

9.1

 

Mounting/start-up 

The variable-area flowmeter must be installed perpendicularly (direction of flow from the bottom to the top). 
The size of the product line to be connected must be identical to the size of the device connection.  
Please make sure that there is adequate clearance from parts that might cause magnetic interferences, such as 
solenoid valves and ferromagnetic components like steel brackets/supports. For two adjacently mounted devices 
we recommend a the minimum lateral distance between them of 300 mm. The devices can be mounted closer 
together if they are vertically offset by one device length. The minimum lateral clearance for interfering steel 
parts should be 200 mm. In case of doubt, check the interference by moving the device back and forth in the 
selected distance by about 200 mm and testing whether the pointer position changes. 
Select the mounting location so as to enable a reliable reading of the scale values.  Please take note as well of the 
space requirement for any possible disassembly of the device.   
To achieve stress-free mounting, the flanges of the pipe must be aligned parallel to each other. Please check 
whether the pipe is adequately stable to rule out the possibility of vibration or swinging. (Do not use steel 
mounting parts on the device.) 
If there is risk of dirt or solid matter penetrating the process pipes, flush them beforehand so that these materials 
do not get caught in the device.  Ferromagnetic solid matter such as spatter can lead to the breakdown of the 
device.  If these materials are still present during normal operating conditions, mount a magnetic filter 
(accessory) in front of the device. 
When using gases, slowly let the inlet pressure rise to avoid pressure surges. 
Basically, to prevent the float from shooting upwards during flow start-up, the use  of solenoid valves should be 
avoided. 
KDS-S, BGK: Avoid installing a valve directly in front of the device. 
9.1.1

 

Device setting 

The measuring equipment is delivered ready for operation according to your order specifications.  The limit 
transducers
 are set to the desired values.  If you have not submitted any requirements, the basic settings are the 
following: 
1 contact device: - Minimum contact switching point at 10% of descending flow (damped/closed-circuit principle) 
2 contact devices: Minimum contact switching point at 10% of descending flow and maximum contact switching 
point at 90% of ascending flow 

Содержание Heinrichs BGK

Страница 1: ...nik GmbH Operating manual KDS BGK January 2021 Variable Area Flowmeters KDS BGK Operating Instructions KDS K C S KDS R BGK KDS E analogue output Betriebsanleitung bitte durchlesen und an einem sichere...

Страница 2: ...design 8 6 Input 9 6 1 Measured variable 9 6 2 Measuring range 9 6 3 Regulator range 9 6 4 Measuring range table 9 7 Electrical output 10 7 1 Limit switches 10 7 2 Analogue output 10 8 Measuring accu...

Страница 3: ...E2 22 13 Indicator unit 23 14 Auxiliary power for limit switch 23 15 Analog output 4 20 mA 23 15 1 Measuring principle 23 15 2 System configuration 23 15 3 Measured variable 23 15 4 Parameters 23 15...

Страница 4: ...lations start up maintenance and operation They must read and understand the operating manual and follow its instructions Follow the conditions and provisions applicable in your country Heinrichs Mess...

Страница 5: ...ory issued packaging and the transport protection 2 4 Returning devices for repair and service Note In accordance with the applicable German waste disposal legislation the owner client is responsible...

Страница 6: ...Product type KDS K C S and BGK Miniature flowmeter in full metal design based on the float principle KDS R Miniature flowmeter in all metal design based on the float principle with differential pressu...

Страница 7: ...sing monitoring adjusting and controlling them They show the current flow quantity in volume or mass per unit in time 4 2 KDS R The KDS R flowmeter can be used for stabilizing set flow quantities of l...

Страница 8: ...anges either on the input side or the output side the built in diaphragm valve causes a pressure compensation that keeps the set flow rate constant Important The regulator can only adjust pressure flu...

Страница 9: ...ure loss H2O mbar KDS R Pressure loss H2O mbar KDS K C Pressure loss H2O mbar KDS S BGK KDS K C R KDS S BGK A 0 1 1 0 3 30 350 6 6 B 0 25 2 5 5 50 350 7 7 5 C 0 6 6 0 18 180 350 7 7 5 D 1 0 10 30 300...

Страница 10: ...Pepperl Fuchs Type NJ2 11 SN Connection M12 plug Ex Marking PTB 00 ATEX 2049 X II 2G Ex ia IIC T6 T4 Gb II 2D Ex ia IIIC T135 C Db 7 2 Analogue output 4 20 mA current loop passive Manufacturer Heinric...

Страница 11: ...ion for KDS R 5 0 of URV Repeatability 1 0 of URV KDS R regulator 1 7 of URV Influence of ambient temperature none 8 1 Influence of fluid temperature Deviations in fluid temperature from the temperatu...

Страница 12: ...enable a reliable reading of the scale values Please take note as well of the space requirement for any possible disassembly of the device To achieve stress free mounting the flanges of the pipe must...

Страница 13: ...ons Class C according to IEC 654 Part 1 Degree of protection Acc To DIN EN 60529 IP 65 t Shock resistance vibration resistance To prevent damagehe meter should be protected from extreme shocks and vib...

Страница 14: ...o PN 40 The maximum unilateral pressure resistance of the diaphragm is 7 bar Inlet and outlet sections Inlet and outlet sections are not required for a linear flow profile of the fluid For a non linea...

Страница 15: ...vertically movable float All metallic wetted parts are made of 1 4404 316L stainless steel The measured values are transmitted magnetically to the analog indicator unit mounted on the device KDS K pr...

Страница 16: ...Heinrichs Messtechnik GmbH Page 16 Betriebsanleitung KDS BGK 10 2 KDS S Process connections vertically at the top at the bottom mounting dimension 180 mm NPT F connection...

Страница 17: ...on The unit consists of a variable area flowmeter with a setting valve and a differential pressure regulator mounted on the meter The measured value is displayed on the scale of the analog indicator u...

Страница 18: ...Heinrichs Messtechnik GmbH Page 18 Betriebsanleitung KDS BGK 10 4 KDS R with NPT F connection as inlet pressure regulator version...

Страница 19: ...nik GmbH Operating Manual KDS BGK Page 19 10 5 BGK with flange connection 10 6 KDS BGK with M12 plug connection All KDS BGK devices equipped with electrical contacts are connected via an M12 plug sock...

Страница 20: ...steel Valve sealing KDS K C R PTFE Tube sealing KDS S Viton Regulator diaphragm KDS R Viton PTFE 10 9 Process connection KDS NPT F Special connections Ermeto 6 8 10 12 15 mm Swagelok 6 8 10 12 mm G M...

Страница 21: ...5 6 5 N Switch K1 L to K1 L Pin2 Namur Switch 12 1 2 M12 standard configuration 2 x inductive sensors NJ 1 5 6 5 N Electrical Output Function KEI 2 NJ 1 5 6 5 N Switch K1 L Pin1 to output K1 L Pin2 S...

Страница 22: ...2 6 5M 30 E2 Electrical Output Funktion KEI 1 NBB 2 6 5M 30 E2 K1 L Pin 1 K1 L Pin 3 K1 output Pin 4 3 wire PNP 12 1 5 M12 standard configuration 2 x inductive sensors NBB 2 6 5M 30 E2 Electrical Out...

Страница 23: ...passive 4 to 20 mA loop powered analogue output Therefore the voltage at terminals plug must never fall short of 14 V DC The 4 to 20 mA output is linearized and corresponds to the indication scale Th...

Страница 24: ...rking may be operated within explosive atmospheres areas Ambient temp Process Temp Marking Zone Description 25 C 70 C 25 C 100 C II 2G Ex h IIC T6 T5 Gb II 2D Ex h IIIC T85 C T100 C Db 1 With switch 2...

Страница 25: ...uction for the measuring tube 16 2 Electrostatic charge of non conductive parts In hazardous areas of protection class IIC pay attention to the risk of the electrostatic charge in dangerous quantities...

Страница 26: ...ed regularly to avoid deposits exceeding 5 mm 16 6 With integrated electrical sensors When the electrical sensors are installed the device becomes an electrical assembly and receives the additional ma...

Страница 27: ...t temperature Tamb 40 C Max fluid temperature Tm 90 C Temperature class T4 T excessive Temperature Ta Ambient temperature of the limit switches T Tm Tamb 90 C 40 C 50 C Ta T 0 4 Tamb 50 C 0 4 40 C 60...

Страница 28: ...55011 Group 1 Class B NAMUR recommendation NE 21 Others EN 60529 Degrees of protection through housing IP code EN 61010 1 Safety requirements for electrical measuring control and laboratory devices E...

Страница 29: ...remove the cover and move the pointer if the pointer can be moved easily the float cannot move If the pointer is unable to move further send the device to the head office for servicing The float is st...

Страница 30: ...a health hazard environmentally hazardous we have performed the following steps Checked all cavities in the device to ensure that they are free of fluid residues Washed and neutralized all cavities i...

Страница 31: ...Heinrichs Messtechnik GmbH Operating Manual KDS BGK Page 31 23 Declaration of conformity...

Страница 32: ...Heinrichs Messtechnik GmbH Page 32 Betriebsanleitung KDS BGK...

Страница 33: ...Heinrichs Messtechnik GmbH Operating Manual KDS BGK Page 33...

Отзывы: