Measurement
Measuring the conductivity (Cond)
Operation
Measuring
mode
Pressing meas accesses the measuring mode from all func-
tions. In the measuring mode, the main display indicates
the measured variable and the secondary display shows
the temperature.
Note
If the Portamess® 913 (X) Cond is connected to a PC and
measurements are taken in a grounded liquid, measuring
errors may result.
The main display indicates the measured conductivity, the
secondary display shows the temperature.
Temperature
compensation
The meter offers various temperature compensation
methods. With meas and
▲
or
▼
, the temperature
compensation method can be selected and set:
(
tc OFF
) No temperature compensation
(
tc nLF
) Temperature compensation with non-linear charac-
teristic to EN 27088 for natural water and ultrapure water
(reference temperature 25 °C). In the secondary display
tc
also appears.
(
tc 0.01 – 9.99 %/°C
) Temperature compensation with lin-
ear characteristic and definable temperature coefficients
(reference temperature 25 °C). In the secondary display
tc
also appears.
Note
When you have selected temperature compensation
with linear characteristic, you can only exit this function
or select the nonlinear function when the temperature
coefficient has been set to 0.00.