Configuration
Operation
The following basic settings can be changed in the
configuration:
Function:
Cond
(conductivity),
SAL
(salinity) or
tdS
(Total
Dissolved Solids or evaporation residue)
• Calibration by entering the cell constant (
AutCAL Off
)
or calibration with calibration solution (
AutCAL On
)
• Automatic meter switch-off after 1 hour or 12 hours
• Interface: Printer output On/Off, baud rate
• Temperature display °C or °F
• Date and time format 24 hours and day, month, year
or 12 hours (a.m./p.m.) and month, day, year
+
To activate the configuration, hold down cal with the
meter switched off and then press on/off.
The menu items of the configuration menu are worked
through in sequence. Use
▲
and
▼
to change the
setting of the respective menu item. STO saves the
parameters and switches to the next menu item.
Pressing meas exits the configuration menu at any
time. The value last displayed and possibly changed will
then not be saved.
Function
Select the measuring function:
Cond
(conductivity),
SAL
(salinity) or
tdS
(Total Dissolved Solids or evaporation
residue).
Automatic
or manual
calibration
Select whether you wish to adapt the sensor by directly
entering the cell constant or by calibrating with a cali-
bration solution and automatic drift check.
(Default setting: Direct entry of the cell constant
(
AutCAL OFF
))
Direct entry of the cell constant (
AutCAL OFF
) from
0.010 cm
–1
to 199.9 cm
–1
. (Default setting 0.475 cm
–1
)