Calibration by direct entry of the cell constant (
AutCAL OFF
)
Operation
SE 202 and
SE 204 sensors
For the conductivity sensor models SE 202 and SE 204, the
cell constant is specified with a tolerance of 2 % and 1.5 %.
This cell constant is entered and stored in the calibration
mode (
AutCAL OFF
). An additional calibration with calibra-
tion solutions is not necessary.
SE 202 sensor: c = 0.1 cm
–1
SE 204 sensor: c = 0.475 cm
–1
Press cal to activate calibration. The cell constant deter-
mined or set during the last calibration is displayed.
Pressing meas exits calibration again.
Use
▲
and
▼
to set the cell constant of the sensor and
confirm by pressing cal. The meter will then switch back to
the measuring mode.