In conjunction with the Unical 9000 probe controller, you can take the required compressed air
from the Aux2 valve of Unical 9000, limit the maximum pressure through a pressure reducer
(accessory ZU 0670/2), and automatically perform a preventive daily condensate drain using a
special program.
Possible Program Flow:
01 Probe in MEASURE
{Make sure that probe is in PROCESS position}
02 Aux medium On 0060s
{Pressurize immersion tube for 1 min}
03 Aux medium Off 0015s
{Deaerate immersion tube, wait for 15 s}
04 Rinse water On 0010s
{Refill ceramic slide with water}
05 Rinse water Off 0000s
06 Program end
Automatic Condensate Drain
Ceramat® WA 160(X) Retractable Fitting
Menu
Display
Parameter setting
Condensate drain
Parameter setting, activating
From the measuring mode:
Press
menu
key to select menu.
Select parameter setting using arrow keys,
confirm by pressing
enter
.
Passcode: 1989 (default)
Parameter setting
Select “Probe control”.
Icons to assign the measured values displayed:
specifies module slot I
specifies module slot II
Probe controller
Select “Fixed interval” from the “Time control”
pull-down menu.
Then open the “Fixed interval” menu.
Fixed interval
Select a new user program (e.g. User 2) from the
"Program" pull-down menu.
Menu selection
Return to meas
Select:
[enter]
Lingua
System control
Back
25.6°C
Module BASE 3400-021
Module PHU 3400-110
7.05 pH
Probe controller
Module FRONT 3400-011
OK
Parameter setting (Administrator)
25.6°C
7.05 pH
Probe control (Administrator)
Controller
Back
Fixed interval
Program flow
Time control
Cal preset values
Fixed interval
Installation
Unical 9000
Block
Fixed interval (Administrator)
Interval
OK
25.6°C
Program
7.05 pH
Interval
Program
Interval
Program
Clean
Cal 2point
Cal 1point
Parking
User 2
Off
User 3
Back