Information on Installation
Ceramat® WA 160(X) Retractable Fitting
For operation of the Ceramat®, you must connect control air, rinsing or calibration media, and the
electrical check-back signal for indicating the probe position.
There are two possible ways to do so.
When the Ceramat is operated with a Unical® or Uniclean® electro-pneumatic controller and the
Protos® measuring system, the cables and tubings for air pressure, rinsing/calibration media and
check-back are combined in a single hose with just one plug connection (multiplug). This hose is
referred to as media connection.
This media connection is installed on the Ceramat® together with the outlet hose.
When you do not use a probe controller (Unical® or Uniclean® and Protos® measuring system), you
can connect the supply lines for control air, rinsing/calibration media and electrical check-back to
the Ceramat® with a free hose connection via the ZU 0631 standard-media interface.
Connection for operation with
Unical®/Uniclean® and Protos®
Connection for operation with
ZU 0631 standard media interface