35
Menu item
Action
Choices
TAG
Select ON or OFF using
keys.
Press enter to confirm.
When switched on, the entry for
“TAG” in the Memosens sensor
is compared to the entry in the
analyzer.
If the entries differ, a message
will be generated.
GROUP
Select ON or OFF using
keys.
Press enter to confirm.
Function as described above
Sensor Verification (TAG, GROUP)
When Memosens sensors are calibrated in the lab, it is often useful and sometimes even mandatory that
these sensors will be operated again at the same measuring points or at a defined group of measuring
points. To ensure this, you can save the respective measuring point (TAG) or group of measuring points
(GROUP) in the sensor. TAG and GROUP can be specified by the calibration tool or automatically entered
by the transmitter. When connecting an MS sensor to the transmitter, it can be checked if the sensor con-
tains the correct TAG or belongs to the correct GROUP. If not, a message will be generated and Sensoface
gets “sad”. The “sad” Sensoface icon can also be signaled by a 22 mA error current. Sensor verification can
be switched on in the Configuration in two steps as TAG and GROUP if required.
When no measuring point or group of measuring points is saved in the sensor, e.g., when using a new
sensor, Stratos enters its own TAG and GROUP. When sensor verification is switched off, Stratos always
enters its own measuring point and group. A possibly existing TAG/GROUP will be overwritten.
Configuration